Real time continuous monitoring of injection pressure at the needle tip is better than ultrasound in early detecting intraneural injection
Regional Anesthesia & Pain Medicine · 2026-04 · 51 (4) · 460-464
DOI: 10.1136/rapm-2024-106086 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Regional Anesthesia & Pain Medicine, online erschienen am 11.12.2024. Verfasst von Roberto Dossi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Anesthesia, Ospedale Regionale di Bellinzona e Valli Bellinzona, Bellinzona, Switzerland roberto.dossi87@gmail.com. Hauptthemen laut MEDLINE: Needles, Ultrasonography, Interventional, Nerve Block, Peripheral Nerves. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 28 Literaturangaben.
Abstract (Originalsprache)
Introduction
During peripheral nerve blocks, inadvertent intraneural injection is not infrequent. Recently, with sonographic nerve swelling, it has been shown that intraneural injection can be detected as early as 0.4 mL. A new method based on injection pressure monitoring at the needle tip, the real pressure in the tissues, is compared with sonographic performance in early detection of intraneural injection.
Methods
This cadaver study was conducted on the upper limb of fresh cadavers. Senior anesthesiologists performed ultrasound guided intraneural injections of the nerves using a modified Visioplex needle with a fiberoptic sensor embedded. 12 injections were performed, two on each nerve (radial, ulnar, and median) for each anatomical preparation. 3 mL of saline methylene blue mix was injected at 10 mL/min. Recordings of ultrasound screen and pressure curves were blindly analyzed to evaluate nerve swelling and injection pressures.
Results
An immediate increase in injection pressure from time zero was observed in all 12 intraneural injections, with similar trends. After only 0.2 mL, intraneural injection was clearly identified as the pressure curves had already risen to an average of 120 mm Hg (SD 5). Nerve swelling was observed in 10 of 12 injections (83%). This was evident after an average of 1.2 mL (SD 0.5). Following injections, dissection of specimens confirmed intraneural spread in all cases.
Conclusion
Pressure monitoring at the needle tip consistently showed a pressure rise from the onset of injection and proved to be a more sensitive and earlier indicator of intraneural injection than sonographic nerve swelling.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Anesthesia, Ospedale Regionale di Bellinzona e Valli Bellinzona, Bellinzona, Switzerland roberto.dossi87@gmail.com.
- Clinica Sant'Anna, Sorengo, Switzerland.
- Anesthesiology and Critical Care department, Hopital Lapeyronie, Montpellier, France.
- Anesthesia, Ospedale Regionale di Bellinzona e Valli Bellinzona, Bellinzona, Switzerland.
- Faculty of Biomedical Sciences, University of Lugano (USI), Lugano, Switzerland.
ORCID-Profile: Roberto Dossi, Christian Quadri, Xavier Capdevila, Andrea Saporito
Zitieren
APA
Dossi, R., Quadri, C., Capdevila, X., & Saporito, A. (2026). Real time continuous monitoring of injection pressure at the needle tip is better than ultrasound in early detecting intraneural injection. Regional Anesthesia & Pain Medicine, 51(4), 460-464. https://doi.org/10.1136/rapm-2024-106086
Vancouver
Dossi R, Quadri C, Capdevila X, Saporito A. Real time continuous monitoring of injection pressure at the needle tip is better than ultrasound in early detecting intraneural injection. Regional Anesthesia & Pain Medicine. 2026;51(4):460-464. doi:10.1136/rapm-2024-106086
BibTeX
@article{Dossi2026,
author = {Dossi, Roberto and Quadri, Christian and Capdevila, Xavier and Saporito, Andrea},
title = {Real time continuous monitoring of injection pressure at the needle tip is better than ultrasound in early detecting intraneural injection},
journal = {Regional Anesthesia & Pain Medicine},
year = {2026},
volume = {51},
number = {4},
pages = {460--464},
doi = {10.1136/rapm-2024-106086}
}
EUVIMED