Noninvasive measurement of glucose and lactate in thermoregulatory sweat in neonates and adults

Xinyue Liu iD, Sally A N Gowers iD, Batia Gourin, Zoe McClure, Sarah Jane Dado, Theresa Abraham, Jayanta Banerjee iD, Martyn G Boutelle iD

Science Advances · 2026-07-17 · 12 (29)

DOI: 10.1126/sciadv.aec4166 ↗

At a glance

Journal article in Science Advances, published on 2026-07-17. Written by Xinyue Liu and 7 co-authors. Institution: Department of Bioengineering, Imperial College London, UK. Major MEDLINE topics: Sweat, Lactic Acid, Glucose, Body Temperature Regulation. Cited 1 times so far (Crossref); 60 references. The full text is freely available in PubMed Central. Funded by: Engineering and Physical Sciences Research Council, NIHR Imperial Biomedical Research Centre.

SantraukaSubjectsBibliografiniai duomenysCiteRelated studies

Abstract (original language)

Premature neonates are particularly vulnerable and therefore require rigorous monitoring of analytes such as glucose and lactate. Traditionally, this is achieved through invasive, intermittent blood sampling, which is painful and may cause potential complications. To address this issue, we have developed a noninvasive perfusable skin sampling patch designed to collect analytes from low-volume thermoregulatory sweat without the need for external sweat induction. This can be combined with microfluidic biosensor analysis technologies to detect skin glucose and lactate levels. We were able to demonstrate the potential of this technology for noninvasive measurement of glucose and lactate in premature neonates. Our results showed that skin glucose levels correlated well with time-matched blood glucose measurements. Lactate skin measurements did not correlate with blood measurements, indicating a more complex relationship. In addition, we successfully measured skin glucose levels in healthy adults, which correlated with blood glucose levels for each individual. In contrast with the neonate measurements, the relationship between blood and skin glucose varies between individual adults. We successfully carried out proof-of-concept experiments to detect glucose in real-time in both neonates and in healthy adults by connecting the sampling patch to a continuous analysis system. These results provide the groundwork for subsequent large-scale clinical trials intended to validate the system reliability and clinical applicability within health care contexts, particularly neonatal intensive care units.

Subjects

MeSH terms (MEDLINE)

SweatLactic AcidGlucoseBody Temperature RegulationHumansInfant, NewbornAdultBiosensing TechniquesBlood GlucoseSkinFemaleMale

Substances

Lactic AcidGlucoseBlood Glucose

Authors and institutions

  • Department of Bioengineering, Imperial College London, UK.
  • Department of Neonatology, Women and Children's Division, Imperial College Healthcare Trust, UK.

ORCID profiles: Xinyue Liu, Sally A N Gowers, Jayanta Banerjee, Martyn G Boutelle

Funding

  • Engineering and Physical Sciences Research Council
  • NIHR Imperial Biomedical Research Centre

Cite

APA

Liu, X., Gowers, S. A. N., Gourin, B., McClure, Z., Dado, S. J., Abraham, T., Banerjee, J., & Boutelle, M. G. (2026). Noninvasive measurement of glucose and lactate in thermoregulatory sweat in neonates and adults. Science Advances, 12(29). https://doi.org/10.1126/sciadv.aec4166

Vancouver

Liu X, Gowers SAN, Gourin B, McClure Z, Dado SJ, Abraham T, et al. Noninvasive measurement of glucose and lactate in thermoregulatory sweat in neonates and adults. Science Advances. 2026;12(29). doi:10.1126/sciadv.aec4166

BibTeX
@article{Liu2026,
  author = {Liu, Xinyue and Gowers, Sally A N and Gourin, Batia and McClure, Zoe and Dado, Sarah Jane and Abraham, Theresa and Banerjee, Jayanta and Boutelle, Martyn G},
  title = {Noninvasive measurement of glucose and lactate in thermoregulatory sweat in neonates and adults},
  journal = {Science Advances},
  year = {2026},
  volume = {12},
  number = {29},
  doi = {10.1126/sciadv.aec4166}
}
Start for free →