A Multichamber Pulsating-Flow Device With Optimized Spatial Shear Stress and Pressure for Endothelial Cell Testing
Journal of Biomechanical Engineering · 2025
DOI: 10.1115/1.4066800 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomechanical Engineering, 2025. Verfasst von Obed A Campos und 4 weiteren Autor:innen. Beteiligte Einrichtung: Mechanical and Aerospace Engineering, University of California San Diego, La Jolla, CA 92093-0411. Hauptthemen laut MEDLINE: Stress, Mechanical, Pressure, Endothelial Cells. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Design and analysis are presented for a new device to test the response of endothelial cells to the simultaneous action of cyclic shear stresses and pressure fluctuations. The design consists of four pulsatile-flow chambers connected in series, where shear stress is identical in all four chambers and pressure amplitude decreases in successive chambers. Each flow chamber is bounded above and below by two parallel plates separated by a small gap. The design of the chamber planform must ensure that cells within the testing region experience spatially uniform time-periodic shear stress. For conditions typically encountered in applications, the viscous unsteady flow exhibits order-unity values of the associated Womersley number. The corresponding solution to the unsteady lubrication problem, with general nonsinusoidal flowrate, is formulated in terms of a stream function satisfying Laplace's equation, which can be integrated numerically to determine the spatial distribution of shear stresses for chambers of general planform. The results are used to optimize the design of a device with a hexagonal planform. Accompanying experiments using particle tracking velocimetry (PTV) in a fabricated chamber were conducted to validate theoretical predictions. Pressure readings indicate that intrachamber pressure variations associated with viscous pressure losses and acoustic fluctuations are relatively small, so that all cells in a given testing region experience nearly equal pressure forces.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Mechanical and Aerospace Engineering, University of California San Diego, La Jolla, CA 92093-0411.
- Department of Pediatrics, University of San Francisco, San Francisco, CA 94158; Cardiovascular Research Institute, University of San Francisco, San Francisco, CA 94158.
Förderung
- NHLBI NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Campos, O. A., Garcia-Herreros, A., Sánchez, A. L., Fineman, J. R., & Pawlak, G. (2025). A Multichamber Pulsating-Flow Device With Optimized Spatial Shear Stress and Pressure for Endothelial Cell Testing. Journal of Biomechanical Engineering. https://doi.org/10.1115/1.4066800
Vancouver
Campos OA, Garcia-Herreros A, Sánchez AL, Fineman JR, Pawlak G. A Multichamber Pulsating-Flow Device With Optimized Spatial Shear Stress and Pressure for Endothelial Cell Testing. Journal of Biomechanical Engineering. 2025. doi:10.1115/1.4066800
BibTeX
@article{Campos2025,
author = {Campos, Obed A and Garcia-Herreros, Antoni and Sánchez, Antonio L and Fineman, Jeffrey R and Pawlak, Geno},
title = {A Multichamber Pulsating-Flow Device With Optimized Spatial Shear Stress and Pressure for Endothelial Cell Testing},
journal = {Journal of Biomechanical Engineering},
year = {2025},
doi = {10.1115/1.4066800}
}
EUVIMED