Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious Cats
Veterinary Ophthalmology · 2027-01 · 30 (1)
DOI: 10.1111/vop.70304 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, online erschienen am 18.11.2026. Verfasst von Mehmet Nur Çetin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Veterinary Medicine, Department of Surgery, Burdur Mehmet Akif Ersoy University, Burdur, Turkey. Die Arbeit nennt 18 Literaturangaben.
Abstract (Originalsprache)
Rebound tonometry is widely used for measuring intraocular pressure (IOP) in cats; however, differences in clinical tonometer measurement configuration may contribute to variability in recorded values. This prospective repeated‐measures study compared IOP measurements obtained using horizontal and vertical Tonovet Plus clinical measurement configurations in 32 client‐owned, systemically and ophthalmologically normal conscious cats. Intraocular pressure was measured in the right eye while the probe was directed toward the center of the cornea and positioned 4–8 mm from the corneal surface. The order of horizontal and vertical configurations was randomized for each cat. Measurements were obtained by one experienced veterinary ophthalmologist and a trained student operator group. In the primary linear mixed‐effects analysis accounting for repeated measurements within cats, horizontal configuration was associated with an estimated 2.41‐mmHg higher IOP than vertical configuration ( β = 2.41 mmHg; 95% CI, 1.39–3.44; p < 0.001). The configuration‐associated difference was also observed separately for the experienced instructor (mean difference, 2.81 mmHg; 95% CI, 1.91–3.71; p < 0.001) and across the trained student operator group (2.00 mmHg; 95% CI, 0.81–3.19; p = 0.002). These findings indicate that horizontal and vertical Tonovet Plus clinical measurement configurations may yield different IOP values in conscious cats. As configuration changes may affect instrument handling, restraint, head positioning, and animal responses, the difference cannot be attributed solely to tonometer orientation or mechanics. Consistent configuration is therefore recommended when comparing serial IOP measurements in cats. Further controlled studies are needed to distinguish mechanical, ergonomic, and other examination‐related mechanisms.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Veterinary Medicine, Department of Surgery, Burdur Mehmet Akif Ersoy University, Burdur, Turkey.
- Cerrahpasa Veterinary Faculty Institute of Health Sciences, Istanbul University, Istanbul, Turkey.
ORCID-Profile: Mehmet Nur Çetin, Yusuf Sinan Şirin, Batuhan Neyse
Zitieren
APA
Çetin, M. N., Şirin, Y. S., Polat, D., Neyse, B., & Yanmaz, L. E. (2027). Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious Cats. Veterinary Ophthalmology, 30(1). https://doi.org/10.1111/vop.70304
Vancouver
Çetin MN, Şirin YS, Polat D, Neyse B, Yanmaz LE. Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious Cats. Veterinary Ophthalmology. 2027;30(1). doi:10.1111/vop.70304
BibTeX
@article{etin2027,
author = {Çetin, Mehmet Nur and Şirin, Yusuf Sinan and Polat, Doğukan and Neyse, Batuhan and Yanmaz, Latif Emrah},
title = {Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious Cats},
journal = {Veterinary Ophthalmology},
year = {2027},
volume = {30},
number = {1},
doi = {10.1111/vop.70304}
}
EUVIMED