Measurement of Airway Length in Neonates Using Fiberoptic Bronchoscopy
Pediatric Anesthesia · 2026
DOI: 10.1111/pan.70028 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Anesthesia, 2026. Verfasst von Vanya Chugh und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesia and Intensive Care, Vardhman and Mahavir Medical College and Safdarjung Hospital, New Delhi, India. Hauptthemen laut MEDLINE: Bronchoscopy, Fiber Optic Technology, Airway Management, Trachea. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben.
Abstract (Originalsprache)
Introduction
Optimal positioning of the endotracheal tube is a crucial step in the airway management of neonates. Short airway length creates a narrow margin of safety and thus a higher chance of ETT malpositioning. Published literature reports airway length of 5–7.5 cm, which is primarily based on autopsy and radiologic studies. We decided to measure the airway length in neonates undergoing surgeries using flexible fiberoptic bronchoscopy (FFOB). Methodology Sixty full‐term neonates, scheduled for surgery under general anesthesia with endotracheal intubation, were included in the study. Neonates with airway anomalies and anticipated difficult airways were excluded. Airway length was measured using FFOB.
Results
Sixty neonates were included in the study. The mean airway length ± SD was 3.85 ± 0.71 cm. The mean ± SD length in males was 3.87 ± 0.63 cm, and in females was 3.82 ± 0.86 cm. The Spearman rank correlation coefficient of airway length (cm) with age was 0.139, and for weight was 0.130.
Conclusion
Airway length is much smaller in neonates as compared to that reported in the literature. Airway dimensions need reassessment in this population in order to improve safety and effectiveness of neonatal intubation practices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesia and Intensive Care, Vardhman and Mahavir Medical College and Safdarjung Hospital, New Delhi, India.
ORCID-Profile: Vanya Chugh
Zitieren
APA
Chugh, V., Kashyap, R., & Bamba, C. (2026). Measurement of Airway Length in Neonates Using Fiberoptic Bronchoscopy. Pediatric Anesthesia. https://doi.org/10.1111/pan.70028
Vancouver
Chugh V, Kashyap R, Bamba C. Measurement of Airway Length in Neonates Using Fiberoptic Bronchoscopy. Pediatric Anesthesia. 2026. doi:10.1111/pan.70028
BibTeX
@article{Chugh2026,
author = {Chugh, Vanya and Kashyap, Rohit and Bamba, Charu},
title = {Measurement of Airway Length in Neonates Using Fiberoptic Bronchoscopy},
journal = {Pediatric Anesthesia},
year = {2026},
doi = {10.1111/pan.70028}
}
EUVIMED