Evaluation of Lead Properties Affecting Complete Radiological Success in Transvenous Lead Extraction
Pacing and Clinical Electrophysiology · 2026-07-10
DOI: 10.1111/pace.70342 ↗At a glance
Journal article in Pacing and Clinical Electrophysiology, published online on 2026-07-10. Written by Murat Erdem Alp and 6 co-authors. Institution: Başakşehir Çam and Sakura City Hospital, Istanbul, Turkey. 25 references.
Abstract (original language)
Background
The insertion of cardiac implantable electronic devices has become increasingly common, resulting in a growing need for lead extraction, most often due to infection and malfunction. The majority of the research indicated that the lead indwell time was the most important factor in determining the procedural success of extraction. In this study, we examined the factors affecting complete radiological success.
Methods
A total of 192 lead extractions were performed on 102 patients at Başakşehir Çam and Sakura City Hospital. These procedures were compared based on lead indwell time and lead properties, considering the complete radiological success.
Results
Complete radiological success was determined by lead indwell time and the presence of ICD leads. For leads with an indwell time of less than 12 years, the complete radiological success increased more than fivefold. Furthermore, in patients with ICD leads and lead indwell time exceeding 12 years, more complex and additional extraction techniques may be beneficial rather than simple traction alone.
Conclusion
Prolonged lead indwell time and the presence of ICD leads were identified as two key factors affecting both complete radiological success and procedural difficulty.
Subjects
Author keywords
Authors and institutions
- Başakşehir Çam and Sakura City Hospital, Istanbul, Turkey.
- Konya Kulu Region State Hospital, Kulu, Turkey.
ORCID profiles: Murat Erdem Alp
Cite
APA
Alp, M. E., Polat, V., Barutçu, S., Akgün, T., Bostanci Alp, G. İ., Akgün, H., & Özkan, E. (2026). Evaluation of Lead Properties Affecting Complete Radiological Success in Transvenous Lead Extraction. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70342
Vancouver
Alp ME, Polat V, Barutçu S, Akgün T, Bostanci Alp Gİ, Akgün H, et al. Evaluation of Lead Properties Affecting Complete Radiological Success in Transvenous Lead Extraction. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70342
BibTeX
@article{Alp2026,
author = {Alp, Murat Erdem and Polat, Veli and Barutçu, Süleyman and Akgün, Taylan and Bostanci Alp, Güngör İlayda and Akgün, Hüseyin and Özkan, Eyüp},
title = {Evaluation of Lead Properties Affecting Complete Radiological Success in Transvenous Lead Extraction},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70342}
}
EUVIMED