Transvenous lead extraction: Experience of the Tandem approach
Europace · 02.11.2023 · 25 (11)
DOI: 10.1093/europace/euad331 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, online erschienen am 04.11.2023. Verfasst von Zaki Akhtar und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, St George's University Hospital, Blackshaw Road, Tooting, London SW17 0QT, UK. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Pacemaker, Artificial. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Abbott Medical.
Abstract (Originalsprache)
Aims
Transvenous lead extraction (TLE) is important in the management of cardiac implantable electronic devices but carries risk. It is most commonly completed from the superior access, often with ‘bail-out’ support via the femoral approach. Superior and inferior access may be used in tandem, which has been proposed as an advance in safety and efficacy. The aim of this study is to evaluate the safety and efficacy of the Tandem approach.
Method
The ‘Tandem’ procedure entailed grasping of the targeted lead in the right atrium to provide countertraction as a rotational dissecting sheath was advanced over the lead from the subclavian access. Consecutive ‘Tandem’ procedures performed by a single operator between December 2020 and March 2023 in a single large-volume TLE centre were included and compared with the conventional superior approach (control) using 1:1 propensity score matching; patients were statistically matched for demographics.
Results
The Tandem in comparison with the conventional approach extracted leads of much greater dwell time (148.9 ± 79 vs. 108.6 ± 77 months, P < 0.01) in a shorter procedure duration (96 ± 36 vs. 127 ± 67 min, P < 0.01) but requiring more fluoroscopy (16.4 ± 10.9 vs. 10.8 ± 14.9 min, P < 0.01). The Tandem and control groups had similar clinical (100% vs. 94.7%, P = 0.07) and complete (94.8% vs. 92.8%, P = 0.42) success, with comparable minor (4% vs. 6.7%, P = 0.72) and major (0% vs. 4%, P = 0.25) complications; procedural (0% vs. 1.3%, P = 1) and 30-day (1.3% vs. 4%, P = 0.62) mortality were also similar.
Conclusion
The Tandem procedure is as safe and effective as the conventional TLE. It can be applied to leads of a long dwell time with a potentially shorter procedure duration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, St George's University Hospital, Blackshaw Road, Tooting, London SW17 0QT, UK.
- Department of Cardiology, Ain Shams University, Cairo, Egypt.
- Department of Cardiothoracic Surgery, German Heart Centre, Berlin, Germany.
ORCID-Profile: Zaki Akhtar, Christos Kontogiannis, Ahmed I Elbatran, Lisa W M Leung, Christoph T Starck, Mark M Gallagher
Förderung
- Abbott Medical
Zitieren
APA
Akhtar, Z., Kontogiannis, C., Elbatran, A. I., Leung, L. W. M., Starck, C. T., Zuberi, Z., Sohal, M., & Gallagher, M. M. (2023). Transvenous lead extraction: Experience of the Tandem approach. Europace, 25(11). https://doi.org/10.1093/europace/euad331
Vancouver
Akhtar Z, Kontogiannis C, Elbatran AI, Leung LWM, Starck CT, Zuberi Z, et al. Transvenous lead extraction: Experience of the Tandem approach. Europace. 2023;25(11). doi:10.1093/europace/euad331
BibTeX
@article{Akhtar2023,
author = {Akhtar, Zaki and Kontogiannis, Christos and Elbatran, Ahmed I and Leung, Lisa W M and Starck, Christoph T and Zuberi, Zia and Sohal, Manav and Gallagher, Mark M},
title = {Transvenous lead extraction: Experience of the Tandem approach},
journal = {Europace},
year = {2023},
volume = {25},
number = {11},
doi = {10.1093/europace/euad331}
}
EUVIMED