Electrocardiographic Characteristics Associated With Lead Position Variations in Left Bundle Branch Area Pacing
Pacing and Clinical Electrophysiology · 2026-05-28
DOI: 10.1111/pace.70307 ↗At a glance
Journal article in Pacing and Clinical Electrophysiology, published online on 2026-05-28. Written by Naoki Matsumoto and 9 co-authors. Institution: Department of Cardiology, Osaka City General Hospital, Osaka, Japan. 23 references.
Abstract (original language)
Background
Physiology‐based electrocardiographic (ECG) criteria are used to assess left bundle branch area pacing (LBBAP), but the impact of lead implantation site on ECG parameters remains unclear.
Methods
We analyzed 212 patients who underwent successful LBBAP and classified them into high septum (HS, n = 49), mid septum (MS, n = 119), and low septum (LS, n = 44) groups according to the polarity of leads II and III. ECG parameters, including QRS morphology in lead V1, stimulus‐to‐R‐wave peak times in V6 and V1 (sV6RWPT, sV1RWPT), and the V6–V1 interpeak interval, were compared. The proportions meeting proposed physiology‐based ECG criteria for left bundle branch (LBB) capture were also evaluated.
Results
The LS group demonstrated the highest prevalence of dominant R waves in V1 (qR: 38.6%, R: 4.5%, p < 0.01). It had a shorter sV6RWPT than the MS group (59 ± 9 ms, p = 0.011), with no significant difference compared with the HS group. It also demonstrated a longer sV1RWPT (105 ± 11 ms, p = 0.023) and V6–V1 interpeak interval (46 ± 10 ms, p < 0.01). Although no significant differences were observed among the three groups regarding output‐dependent QRS transition or LBB potential, the LS group more frequently met the V6–V1 interpeak interval criterion (>44 ms, 86%, p < 0.01).
Conclusions
In LBBAP, low septal lead placement was associated with prominent R waves in V1, shorter sV6RWPT versus the MS group, and longer V6–V1 interpeak intervals, highlighting the importance of implantation site when interpreting ECG findings.
Subjects
Author keywords
Authors and institutions
- Department of Cardiology, Osaka City General Hospital, Osaka, Japan.
- Department of Cardiovascular Medicine, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan.
ORCID profiles: Kenji Shimeno, Masanori Matsuo, Takanori Yamazaki
Cite
APA
Matsumoto, N., Shimeno, K., Kashima, Y., Kure, Y., Matsuo, M., Sakamoto, Y., Yamazaki, T., Matsumura, Y., Abe, Y., & Fukuda, D. (2026). Electrocardiographic Characteristics Associated With Lead Position Variations in Left Bundle Branch Area Pacing. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70307
Vancouver
Matsumoto N, Shimeno K, Kashima Y, Kure Y, Matsuo M, Sakamoto Y, et al. Electrocardiographic Characteristics Associated With Lead Position Variations in Left Bundle Branch Area Pacing. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70307
BibTeX
@article{Matsumoto2026,
author = {Matsumoto, Naoki and Shimeno, Kenji and Kashima, Yumi and Kure, Yusuke and Matsuo, Masanori and Sakamoto, Yuya and Yamazaki, Takanori and Matsumura, Yoshiki and Abe, Yukio and Fukuda, Daiju},
title = {Electrocardiographic Characteristics Associated With Lead Position Variations in Left Bundle Branch Area Pacing},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70307}
}
EUVIMED