Customized Sheath Modification for Bachmann Bundle Area Pacing With a Stylet‐Driven Lead
Pacing and Clinical Electrophysiology · 2026
DOI: 10.1111/pace.70263 ↗Kurz gefasst
Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2026. Verfasst von Yousaku Okubo und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Hiroshima University Graduate School of Biomedical and Health Sciences, Hiroshima, Japan. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Cardiac Pacing, Artificial, Electrodes, Implanted. Die Arbeit nennt 5 Literaturangaben.
Abstract (Originalsprache)
Bachmann bundle area pacing (BBp) has primarily been performed using lumenless leads delivered through dedicated atrial sheaths. With stylet‐driven leads, stable positioning at the superior interatrial septum can be technically challenging. We report two patients in whom the distal portion of a standard ventricular delivery catheter was shortened and reshaped using controlled heat molding with a heat gun. The customized sheath enabled stable placement of a stylet‐driven lead at the high interatrial septum, resulting in successful BBp with favorable electrical parameters. Customized sheath modification may facilitate BBp using conventional pacing systems. Clinical Trial Registration : This study was not registered in a clinical trial registration, as it represents a retrospective case series and dose not meet the criteria for a prospective interventional clinical trial.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Medicine, Hiroshima University Graduate School of Biomedical and Health Sciences, Hiroshima, Japan.
ORCID-Profile: Yousaku Okubo, Yukiko Nakano
Zitieren
APA
Okubo, Y., Miyamoto, S., Ishida, S., Maeda, J., Ishibashi, N., Okamura, S., Tokuyama, T., Oda, N., & Nakano, Y. (2026). Customized Sheath Modification for Bachmann Bundle Area Pacing With a Stylet‐Driven Lead. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70263
Vancouver
Okubo Y, Miyamoto S, Ishida S, Maeda J, Ishibashi N, Okamura S, et al. Customized Sheath Modification for Bachmann Bundle Area Pacing With a Stylet‐Driven Lead. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70263
BibTeX
@article{Okubo2026,
author = {Okubo, Yousaku and Miyamoto, Shogo and Ishida, Shunsuke and Maeda, Junji and Ishibashi, Naoki and Okamura, Sho and Tokuyama, Takehito and Oda, Noboru and Nakano, Yukiko},
title = {Customized Sheath Modification for Bachmann Bundle Area Pacing With a Stylet‐Driven Lead},