EUVIMEDEUVIMED entdecken
Multizentrische Studie

Procedural performance and safety of a helix locking tool for left bundle branch pacing: a prospective multicenter study

Diego Menéndez, Ermengol Vallès, Álvaro Marco Del Castillo, Haritz Arrizabalaga, Teresa Moraleda, Pablo Elpidio, Guillermo Gutierrez, Eduardo Franco, Cristina Lozano-Granero

Journal of Interventional Cardiac Electrophysiology · 03.09.2026 · 69 (9) · 3095-3103

DOI: 10.1007/s10840-026-02423-6 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, online erschienen am 03.09.2026. Verfasst von Diego Menéndez und 8 weiteren Autor:innen. Beteiligte Einrichtung: Arnau de Vilanova University Hospital, Av. Rovira Roure 80, Lleida, 25198, Spain. demenendez.lleida.ics@gencat.cat. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Bundle-Branch Block, Electrodes, Implanted, Pacemaker, Artificial. Die Arbeit nennt 14 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Helix retraction during septal lead deployment can complicate left bundle branch pacing (LBBP) procedures performed with extendable–retractable helix leads. A helix locking tool (HLT) has been developed to stabilize the helix of a stylet-driven lead during implantation, but clinical performance data remain limited. This study aimed to evaluate the acute safety and procedural performance of this tool in a multicenter setting.

Methods

In this prospective multicenter registry, 87 consecutive patients who underwent LBBP were enrolled in eight centers in Spain. Implantation was performed using a stylet-driven Tendril STS lead and CPS Direct 3D sheath, with the HLT used during septal deployment. The primary endpoint was freedom from acute adverse events related to the HLT. Secondary endpoints included procedural success, lead repositioning, and acute electrical parameters.

Results

LBBP was successfully achieved in 79 of 87 patients (90.8%). Lead repositioning was required in 36 cases (41.4%), with a mean of 2.3 ± 1.3 attempts. No cases of helix retraction were observed. Three intraprocedural lead dislodgements (3.4%) occurred and were managed without clinical consequences. No acute complications related to the HLT were reported. Electrical parameters were compatible with conduction system capture.

Conclusions

In this prospective multicenter registry, left bundle branch capture was achieved in 79 of 87 patients (90.8%). No helix retraction or acute adverse events considered related to the HLT by site investigators were observed during 24-hour follow-up. In the absence of a comparator group, the study cannot determine whether the HLT reduces helix retraction or improves procedural outcomes. Graphical abstract Helix Locking Tool (HLT) for left bundle branch pacing: 90.8% procedural success, zero helix retraction, and no tool-related complications in 87 patients across 8 centers. In a prospective multicenter registry of 87 patients undergoing LBBP with a stylet-driven lead and the HLT, left bundle branch capture was achieved in 79 patients (90.8%). No helix retraction or acute adverse events considered related to the HLT by site investigators were observed during 24-hour follow-up.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Pacing, ArtificialBundle-Branch BlockElectrodes, ImplantedPacemaker, ArtificialHumansFemaleProspective StudiesMaleAgedEquipment DesignTreatment OutcomeSpainRegistriesBundle of His

Stichwörter der Autor:innen

Conduction system pacingHelix locking toolLeft bundle branch pacingProcedural performanceStylet-driven lead

Autor:innen und Institutionen

  • Arnau de Vilanova University Hospital, Av. Rovira Roure 80, Lleida, 25198, Spain. demenendez.lleida.ics@gencat.cat.
  • Hospital del Mar, Barcelona, Spain.
  • 12 de Octubre University Hospital, Madrid, Spain.
  • Basurto University Hospital, Bilbao, Spain.
  • Juan Ramon Jiménez University Hospital, Huelva, Spain.
  • Hospital Clínico de Valladolid, Valladolid, Spain.

Zitieren

APA

Menéndez, D., Vallès, E., Del Castillo, Á. M., Arrizabalaga, H., Moraleda, T., Elpidio, P., Gutierrez, G., Franco, E., & Lozano-Granero, C. (2026). Procedural performance and safety of a helix locking tool for left bundle branch pacing: a prospective multicenter study. Journal of Interventional Cardiac Electrophysiology, 69(9), 3095-3103. https://doi.org/10.1007/s10840-026-02423-6

Vancouver

Menéndez D, Vallès E, Del Castillo ÁM, Arrizabalaga H, Moraleda T, Elpidio P, et al. Procedural performance and safety of a helix locking tool for left bundle branch pacing: a prospective multicenter study. Journal of Interventional Cardiac Electrophysiology. 2026;69(9):3095-3103. doi:10.1007/s10840-026-02423-6

BibTeX
@article{Menndez2026,
  author = {Menéndez, Diego and Vallès, Ermengol and Del Castillo, Álvaro Marco and Arrizabalaga, Haritz and Moraleda, Teresa and Elpidio, Pablo and Gutierrez, Guillermo and Franco, Eduardo and Lozano-Granero, Cristina},
  title = {Procedural performance and safety of a helix locking tool for left bundle branch pacing: a prospective multicenter study},
  journal = {Journal of Interventional Cardiac Electrophysiology},
  year = {2026},
  volume = {69},
  number = {9},
  pages = {3095--3103},
  doi = {10.1007/s10840-026-02423-6}
}
Kostenlos starten →