Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study

Peter Bennett iD, Stephen J Hahn, Stephen Daniels

Pacing and Clinical Electrophysiology · 2026-11 · 49 (11) · 1706-1715

DOI: 10.1111/pace.70256 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, published online on 2026-04-18. Written by Peter Bennett and 2 co-authors. Institution: Cardiac Rhythm Management R&D, Boston Scientific, Clonmel, Ireland. Major MEDLINE topics: Cardiac Pacing, Artificial. 19 references.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Subcutaneous and extravascular pacing systems represent an emerging class of cardiac rhythm management devices that deliver therapy without direct myocardial contact. Despite their clinical potential, the excitability characteristics of subcutaneous stimulation are not well defined.

Methods

Acute preclinical studies were conducted in four swine ( n = 4) using a coil‐to‐can subcutaneous pacing configuration. Monophasic and biphasic pulses ranging from 1 to 10 ms and 1 to 200 mA in amplitude were delivered asynchronously at rates 20–30 bpm above the intrinsic heart rate. Myocardial capture was confirmed via high‐fidelity intra‐aortic pressure. Strength‐duration curves were generated, and rheobase and chronaxie values were derived using Lapicque's model.

Results

All subjects exhibited a hyperbolic strength‐duration profile. Rheobase ranged from 19.54 to 77.58 mA, chronaxie varied from 0.70 to 2.94 ms. An inverse relationship was identified between rheobase and chronaxie, suggesting that effects of electrode geometry, tissue impedance, or lead placement may influence the excitability profile. Biphasic waveforms tended to lower capture thresholds versus monophasic stimuli; this difference did not reach statistical significance ( p = 0.161). Inter‐subject variability was consistent with differences in tissue impedance and electrode‐myocardial distance.

Conclusion

Subcutaneous pacing exhibits classical excitability consistent with Lapicque's theory. Thresholds follow predictable dependencies on pulse width and waveform, indicating opportunities to optimize energy delivery safety margins. These data provide a basis for programming strategies and future device designs targeting lower energy use and extended battery life in subcutaneous and extravascular pacing systems.

Subjects

MeSH terms (MEDLINE)

Cardiac Pacing, ArtificialAnimalsSwineHeart RateFemale

Author keywords

biphasic waveformcardiac electrophysiologycardiac stimulationchronaxiemonophasic waveformpreclinical modelstrength‐durationsubcutaneous pacing

Authors and institutions

  • Cardiac Rhythm Management R&D, Boston Scientific, Clonmel, Ireland.
  • Atlantic Technological University (ATU), Sligo, Ireland.
  • School of Electronic Engineering, Dublin City University, Dublin, Ireland.

ORCID profiles: Peter Bennett

Cite

APA

Bennett, P., Hahn, S. J., & Daniels, S. (2026). Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study. Pacing and Clinical Electrophysiology, 49(11), 1706-1715. https://doi.org/10.1111/pace.70256

Vancouver

Bennett P, Hahn SJ, Daniels S. Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study. Pacing and Clinical Electrophysiology. 2026;49(11):1706-1715. doi:10.1111/pace.70256

BibTeX
@article{Bennett2026,
  author = {Bennett, Peter and Hahn, Stephen J and Daniels, Stephen},
  title = {Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  volume = {49},
  number = {11},
  pages = {1706--1715},
  doi = {10.1111/pace.70256}
}
Start for free →