Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study

Peter Bennett iD, Stephen J Hahn, Stephen Daniels

Pacing and Clinical Electrophysiology · 2026-11 · 49 (11) · 1706-1715

DOI: 10.1111/pace.70256 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 18.04.2026. Verfasst von Peter Bennett und 2 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Rhythm Management R&D, Boston Scientific, Clonmel, Ireland. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial. Die Arbeit nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Subcutaneous and extravascular pacing systems represent an emerging class of cardiac rhythm management devices that deliver therapy without direct myocardial contact. Despite their clinical potential, the excitability characteristics of subcutaneous stimulation are not well defined.

Methods

Acute preclinical studies were conducted in four swine ( n = 4) using a coil‐to‐can subcutaneous pacing configuration. Monophasic and biphasic pulses ranging from 1 to 10 ms and 1 to 200 mA in amplitude were delivered asynchronously at rates 20–30 bpm above the intrinsic heart rate. Myocardial capture was confirmed via high‐fidelity intra‐aortic pressure. Strength‐duration curves were generated, and rheobase and chronaxie values were derived using Lapicque's model.

Results

All subjects exhibited a hyperbolic strength‐duration profile. Rheobase ranged from 19.54 to 77.58 mA, chronaxie varied from 0.70 to 2.94 ms. An inverse relationship was identified between rheobase and chronaxie, suggesting that effects of electrode geometry, tissue impedance, or lead placement may influence the excitability profile. Biphasic waveforms tended to lower capture thresholds versus monophasic stimuli; this difference did not reach statistical significance ( p = 0.161). Inter‐subject variability was consistent with differences in tissue impedance and electrode‐myocardial distance.

Conclusion

Subcutaneous pacing exhibits classical excitability consistent with Lapicque's theory. Thresholds follow predictable dependencies on pulse width and waveform, indicating opportunities to optimize energy delivery safety margins. These data provide a basis for programming strategies and future device designs targeting lower energy use and extended battery life in subcutaneous and extravascular pacing systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Pacing, ArtificialAnimalsSwineHeart RateFemale

Stichwörter der Autor:innen

biphasic waveformcardiac electrophysiologycardiac stimulationchronaxiemonophasic waveformpreclinical modelstrength‐durationsubcutaneous pacing

Autor:innen und Institutionen

  • Cardiac Rhythm Management R&D, Boston Scientific, Clonmel, Ireland.
  • Atlantic Technological University (ATU), Sligo, Ireland.
  • School of Electronic Engineering, Dublin City University, Dublin, Ireland.

ORCID-Profile: Peter Bennett

Zitieren

APA

Bennett, P., Hahn, S. J., & Daniels, S. (2026). Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study. Pacing and Clinical Electrophysiology, 49(11), 1706-1715. https://doi.org/10.1111/pace.70256

Vancouver

Bennett P, Hahn SJ, Daniels S. Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study. Pacing and Clinical Electrophysiology. 2026;49(11):1706-1715. doi:10.1111/pace.70256

BibTeX
@article{Bennett2026,
  author = {Bennett, Peter and Hahn, Stephen J and Daniels, Stephen},
  title = {Strength‐Duration Characterisation of Subcutaneous Pacing: A Preclinical Study},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  volume = {49},
  number = {11},
  pages = {1706--1715},
  doi = {10.1111/pace.70256}
}
Kostenlos starten →