EUVIMEDEUVIMED entdecken
Crossref

Hemolysis After Pulsed Field Ablation With a Variable‐Loop Irrigated Circular Catheter

Iwanari Kawamura iD, Shinsuke Miyazaki, Ryosuke Kato, Ken Takahashi, Miho Negishi, Masaki Honda, Ryo Tateishi, Kentaro Goto, Takuro Nishimura, Kazuya Yamao, Kentaro Takahashi, Kensuke Ihara iD und 3 weitere

Pacing and Clinical Electrophysiology · 2026-05 · 49 (5) · 656-659

DOI: 10.1111/pace.70208 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 17.03.2026. Verfasst von Iwanari Kawamura und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan. Hauptthemen laut MEDLINE: Hemolysis, Atrial Fibrillation, Catheter Ablation, Therapeutic Irrigation. Die Arbeit nennt 13 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Hemolysis is a recognized complication of endocardial pulsed field ablation (PFA), and its severity is influenced by factors such as the number of applications, tissue contact, and characteristics of the PFA system. However, most existing data are based on the Pentaspline catheter, and limited information is available regarding hemolysis with other PFA systems. Methods and Results We evaluated hemolysis in seven consecutive patients with paroxysmal atrial fibrillation undergoing PFA using a variable‐loop irrigated circular catheter (VLCC; Varipulse, Biosense Webster) and a dedicated generator (TRUPULSE). All procedures included pulmonary vein and posterior wall isolation, guided by a tissue proximity indicator (TPI). A median of 32 applications was delivered per patient. Plasma‐free hemoglobin levels significantly increased from 0.01 (0.00–0.01) g/dL before the procedure to 0.05 (0.04–0.06) g/dL after the procedure ( p < 0.001). Post‐procedural reductions in hemoglobin levels and increases in bilirubin and lactate dehydrogenase levels were also observed, along with subnormal haptoglobin levels in 50% of patients. Creatine kinase levels increased significantly from 84 (82–146) U/L preoperatively to 323 (272–369) U/L postoperatively ( p = 0.018). No cases of acute kidney injury occurred, and renal function remained stable post‐procedure.

Conclusion

This is the first study to evaluate hemolysis with a VLCC‐based PFA using plasma‐free hemoglobin as a biomarker. These findings underscore the importance of evaluating each PFA system individually to better understand their safety and biophysical effects.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemolysisAtrial FibrillationCatheter AblationTherapeutic IrrigationHumansFemaleMaleIrreversible Electroporation TherapyMiddle AgedTreatment OutcomeAged

Stichwörter der Autor:innen

catheter ablationcomplicationelectroporationhemolysispulsed electrical field

Autor:innen und Institutionen

  • Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan.

ORCID-Profile: Iwanari Kawamura, Kensuke Ihara

Zitieren

APA

Kawamura, I., Miyazaki, S., Kato, R., Takahashi, K., Negishi, M., Honda, M., Tateishi, R., Goto, K., Nishimura, T., Yamao, K., Takahashi, K., Ihara, K., Tao, S., Takigawa, M., & Sasano, T. (2026). Hemolysis After Pulsed Field Ablation With a Variable‐Loop Irrigated Circular Catheter. Pacing and Clinical Electrophysiology, 49(5), 656-659. https://doi.org/10.1111/pace.70208

Vancouver

Kawamura I, Miyazaki S, Kato R, Takahashi K, Negishi M, Honda M, et al. Hemolysis After Pulsed Field Ablation With a Variable‐Loop Irrigated Circular Catheter. Pacing and Clinical Electrophysiology. 2026;49(5):656-659. doi:10.1111/pace.70208

BibTeX
@article{Kawamura2026,
  author = {Kawamura, Iwanari and Miyazaki, Shinsuke and Kato, Ryosuke and Takahashi, Ken and Negishi, Miho and Honda, Masaki and Tateishi, Ryo and Goto, Kentaro and Nishimura, Takuro and Yamao, Kazuya and Takahashi, Kentaro and Ihara, Kensuke and Tao, Susumu and Takigawa, Masateru and Sasano, Tetsuo},
  title = {Hemolysis After Pulsed Field Ablation With a Variable‐Loop Irrigated Circular Catheter},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  volume = {49},
  number = {5},
  pages = {656--659},
  doi = {10.1111/pace.70208}
}
Kostenlos starten →