Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta‐analysis
Pacing and Clinical Electrophysiology · 2024-05 · 47 (5) · 614-625
DOI: 10.1111/pace.14972 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 01.04.2024. Verfasst von Adil Salihu und 13 weiteren Autor:innen. Beteiligte Einrichtung: Service of Cardiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Esophagus. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Introduction
The use of esophageal temperature monitoring (ETM) for the prevention of esophageal injury during atrial fibrillation (AF) ablation is often advocated. However, evidence supporting its use is scarce and controversial. We therefore aimed to review the evidence assessing the efficacy of ETM for the prevention of esophageal injury.
Methods
We performed a meta‐analysis and systematic review of the available literature from inception to December 31, 2022. All studies comparing the use of ETM, versus no ETM, during radiofrequency (RF) AF ablation and which reported the incidence of endoscopically detected esophageal lesions (EDELs) were included.
Results
Eleven studies with a total of 1112 patients undergoing RF AF ablation were identified. Of those patients, 627 were assigned to ETM (56%). The overall incidence of EDELs was 9.8%. The use of ETM during AF ablation was associated with a non significant increase in the incidence of EDELs (12.3% with ETM, vs. 6.6 % without ETM, odds ratio, 1.44, 95%CI, 0.49, 4.22, p = .51, I 2 = 72%). The use of ETM was associated with a significant increase in the energy delivered specifically on the posterior wall compared to patients without ETM (mean power difference: 5.13 Watts, 95% CI, 1.52, 8.74, p = .005).
Conclusions
The use of ETM does not reduce the incidence of EDELs during RF AF ablation. The higher energy delivered on the posterior wall is likely attributable to a false sense of safety that may explain the lack of benefit of ETM. Further randomized controlled trials are needed to provide conclusive results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Service of Cardiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
ORCID-Profile: Adil Salihu
Zitieren
APA
Salihu, A., Lu, H., Maurizi, N., Tzimas, G., Herrera Siklody, C., Le Bloa, M., Domenichini, G., Teres, C., Hugelshofer, S., Monney, P., Pruvot, E., Muller, O., Antiochos, P., & Pascale, P. (2024). Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta‐analysis. Pacing and Clinical Electrophysiology, 47(5), 614-625. https://doi.org/10.1111/pace.14972
Vancouver
Salihu A, Lu H, Maurizi N, Tzimas G, Herrera Siklody C, Le Bloa M, et al. Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta‐analysis. Pacing and Clinical Electrophysiology. 2024;47(5):614-625. doi:10.1111/pace.14972
BibTeX
@article{Salihu2024,
author = {Salihu, Adil and Lu, Henri and Maurizi, Niccolo and Tzimas, Georgios and Herrera Siklody, Claudia and Le Bloa, Mathieu and Domenichini, Giulia and Teres, Cheryl and Hugelshofer, Sarah and Monney, Pierre and Pruvot, Etienne and Muller, Olivier and Antiochos, Panagiotis and Pascale, Patrizio},
title = {Prevention of esophageal lesions during atrial fibrillation catheter ablation using esophageal temperature monitoring: A systematic review and meta‐analysis},
journal = {Pacing and Clinical Electrophysiology},
year = {2024},
volume = {47},
number = {5},
pages = {614--625},
doi = {10.1111/pace.14972}
}
EUVIMED