Role of routine investigations post cardiac devices implants in detecting peri‐procedural complications: A retrospective analysis from a tertiary UK center
Pacing and Clinical Electrophysiology · 2024
DOI: 10.1111/pace.14909 ↗At a glance
Journal article in Pacing and Clinical Electrophysiology, 2024. Written by Christina Menexi and 10 co-authors. Institution: Cardiology Department, Essex Cardiothoracic Center, Basildon, UK. Major MEDLINE topics: Defibrillators, Implantable, Pacemaker, Artificial. 16 references.
Abstract (original language)
Background
Peri‐procedural complications associated with cardiac implantable electronic devices are not uncommon. European Society of Cardiology guidelines recommend device checks of all devices within 72 h of implant. European Heart Rhythm Association expert practical guide on Cardiac implantable electronic devices (CIEDs) recommend that a chest x‐ray (CXR) should be performed within 24 h to rule out pneumothorax and document lead positions. First, the rate of peri‐procedural complications associated with CIED implants at our center, as well as patient and/or procedural‐related factors that are associated with higher rates of complications, is analyzed. Second, the yield of the guideline‐recommended measures in the early detection of peri‐procedural complications is examined. Materials and methods Consecutive de novo transvenous device implants at our center in 2019 were retrospectively analyzed. Patients’ demographics, types and indications for device therapy, procedural reports, device checks, and CXRs were obtained from the hospital electronic records.
Results
A total of 578 patients (Age 74 ± 16 years, 68% male) were included. All patients had routine post‐procedure CXRs and device checks. There were 16 (2.8%) complications; 7 (1.2%) pneumothoraxes, 6 (1%) pericardial effusions, and 3 (0.5%) lead displacements. Procedure time correlated significantly with complications; in uncomplicated cases it was 99 ± 43 min versus 127 ± 50 min in procedures associated with complications ( p = .02).
Conclusions
Routine post CIED implantation CXRs can detect early peri‐procedural complications, while repeat post mobilization device checks has low yield of detection of complications. The only statistically significant predictor of peri‐procedural complications is the duration of the procedure; longer procedures were associated with higher rates of complications.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Cardiology Department, Essex Cardiothoracic Center, Basildon, UK.
- Wessex Cardiothoracic Unit, University Hospital of Southampton NHS Foundation Trust, Southampton, UK.
- School of Medicine, University of Southampton, Southampton, UK.
ORCID profiles: Christina Menexi, Mohamed ElRefai
Cite
APA
Menexi, C., ElRefai, M., Abouelasaad, M., Chua, A. Y. T., Handa, I., Newbery, C., Hoskins, N., Ullah, W., Yue, A., Roberts, P. R., & Paisey, J. (2024). Role of routine investigations post cardiac devices implants in detecting peri‐procedural complications: A retrospective analysis from a tertiary UK center. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14909
Vancouver
Menexi C, ElRefai M, Abouelasaad M, Chua AYT, Handa I, Newbery C, et al. Role of routine investigations post cardiac devices implants in detecting peri‐procedural complications: A retrospective analysis from a tertiary UK center. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14909
BibTeX
@article{Menexi2024,
author = {Menexi, Christina and ElRefai, Mohamed and Abouelasaad, Mohamed and Chua, Anne Y T and Handa, Ishita and Newbery, Clare and Hoskins, Nicola and Ullah, Waqas and Yue, Arthur and Roberts, Paul R and Paisey, John},
title = {Role of routine investigations post cardiac devices implants in detecting peri‐procedural complications: A retrospective analysis from a tertiary UK center},
journal = {Pacing and Clinical Electrophysiology},
year = {2024},
doi = {10.1111/pace.14909}
}
EUVIMED