Left bundle branch area pacing (LBBAP) Auto Threshold algorithms Evaluation for Conduction System Pacing: The LATECS pilot Trial
Pacing and Clinical Electrophysiology · 2023
DOI: 10.1111/pace.14777 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Pacing and Clinical Electrophysiology, 2023. Verfasst von Chiara Ghiglieno und 6 weiteren Autor:innen. Beteiligte Einrichtung: Clinica Cardiologica, Dipartimento Toraco-Cardio-Vascolare, Ospedale Maggiore della Carità, Novara, Italy. Hauptthemen laut MEDLINE: Heart Conduction System, Ventricular Septum. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 30 Literaturangaben.
Abstract (Originalsprache)
Background
Automated threshold measurements (ATM) and output adaptation improved safety and follow‐up of cardiac implantable devices (CIED) in the last years. These algorithms were validated for conventional cardiac pacing; however, they were not suitable for permanent His Pacing. Left bundle branch area pacing (LBBAP) is an emerging technique to obtain physiologic cardiac stimulation; we tried to assess if ATM could be applied to this setting.
Methods
Consecutive patients receiving ATM‐capable CIED and LBBAP in our hospital were enrolled in this prospective, observational trial; they were evaluated 3 months after implant, comparing pacing thresholds manually assessed and obtained via ATM. Subsequent remote follow‐up was carried on when available.
Results
Forty‐five patients were enrolled. ATM for LBBAP lead provided consistent results in all the patients and was therefore activated; mean value of manually obtained LBBAP capture threshold was 0.66 ± 0.19 V versus ATM of 0.64 ± 0.19 V. TOST analysis showed equivalence of the two measures ( p = .66). At subsequent follow‐up (mean follow up 7.7 ± 3.2 months), ATM was effective in assessing pacing thresholds and no clinical adverse event was observed.
Conclusions
ATM algorithms proved equivalent to manual testing in determining capture threshold and were reliably employed in patients receiving LBBAP CIED.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinica Cardiologica, Dipartimento Toraco-Cardio-Vascolare, Ospedale Maggiore della Carità, Novara, Italy.
- Università del Piemonte Orientale Amedeo Avogadro, Novara, Italy.
ORCID-Profile: Chiara Ghiglieno
Zitieren
APA
Ghiglieno, C., Dell'Era, G., Veroli, A., De Vecchi, F., Santagostino, M., Porcellini, S., & Patti, G. (2023). Left bundle branch area pacing (LBBAP) Auto Threshold algorithms Evaluation for Conduction System Pacing: The LATECS pilot Trial. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14777
Vancouver
Ghiglieno C, Dell'Era G, Veroli A, De Vecchi F, Santagostino M, Porcellini S, et al. Left bundle branch area pacing (LBBAP) Auto Threshold algorithms Evaluation for Conduction System Pacing: The LATECS pilot Trial. Pacing and Clinical Electrophysiology. 2023. doi:10.1111/pace.14777
BibTeX
@article{Ghiglieno2023,
author = {Ghiglieno, Chiara and Dell'Era, Gabriele and Veroli, Alessandro and De Vecchi, Federica and Santagostino, Matteo and Porcellini, Stefano and Patti, Giuseppe},
title = {Left bundle branch area pacing (LBBAP) Auto Threshold algorithms Evaluation for Conduction System Pacing: The LATECS pilot Trial},
journal = {Pacing and Clinical Electrophysiology},
year = {2023},
doi = {10.1111/pace.14777}
}
EUVIMED