Unusual atrial pacing induced by type II far‐field p‐wave sensing in a patient with complete atrioventricular block
Pacing and Clinical Electrophysiology · 2024
DOI: 10.1111/pace.14743 ↗At a glance
Fallbericht in Pacing and Clinical Electrophysiology, 2024. Written by Yasunobu Yamagishi and 9 co-authors. Institution: The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan. Major MEDLINE topics: Atrioventricular Block, Pacemaker, Artificial. 8 references.
Abstract (original language)
An 88‐year‐old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12‐lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far‐field sensing of intrinsic atrial waves before atrial events; type II far‐field P‐wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan.
- Department of Heart Rhythm Management, University of Occupational and Environmental Health, Kitakyushu, Japan.
ORCID profiles: Yasunobu Yamagishi, Yasushi Oginosawa
Cite
APA
Yamagishi, Y., Oginosawa, Y., Ishii, N., Nakamura, Y., Yagyu, K., Miyamoto, T., Ohe, H., Kohno, R., Kataoka, M., & Abe, H. (2024). Unusual atrial pacing induced by type II far‐field p‐wave sensing in a patient with complete atrioventricular block. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14743
Vancouver
Yamagishi Y, Oginosawa Y, Ishii N, Nakamura Y, Yagyu K, Miyamoto T, et al. Unusual atrial pacing induced by type II far‐field p‐wave sensing in a patient with complete atrioventricular block. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14743