Systematic review and meta-analysis on the impact on outcomes of device algorithms for minimizing right ventricular pacing
Europace · 03.08.2024 · 26 (8)
DOI: 10.1093/europace/euae212 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Europace, online erschienen am 09.08.2024. Verfasst von Davide Antonio Mei und 10 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology Division, Department of Biomedical, Metabolic and Neural Sciences, University of Modena and Reggio Emilia, Policlinico di Modena, Via del …. Hauptthemen laut MEDLINE: Algorithms, Atrial Fibrillation, Cardiac Pacing, Artificial. Die Arbeit wurde bisher 32-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
Physiological activation of the heart using algorithms to minimize right ventricular pacing (RVPm) may be an effective strategy to reduce adverse events in patients requiring anti-bradycardia therapies. This systematic review and meta-analysis aimed to evaluate current evidence on clinical outcomes for patients treated with RVPm algorithms compared to dual-chamber pacing (DDD). Methods and results We conducted a systematic search of the PubMed database. The predefined endpoints were the occurrence of persistent/permanent atrial fibrillation (PerAF), cardiovascular (CV) hospitalization, all-cause death, and adverse symptoms. We also aimed to explore the differential effects of algorithms in studies enrolling a high percentage of atrioventricular block (AVB) patients. Eight studies (7229 patients) were included in the analysis. Compared to DDD pacing, patients using RVPm algorithms showed a lower risk of PerAF [odds ratio (OR) 0.74, 95% confidence interval (CI) 0.57–0.97] and CV hospitalization (OR 0.77, 95% CI 0.61–0.97). No significant difference was found for all-cause death (OR 1.01, 95% CI 0.78–1.30) or adverse symptoms (OR 1.03, 95% CI 0.81–1.29). No significant interaction was found between the use of the RVPm strategy and studies enrolling a high percentage of AVB patients. The pooled mean RVP percentage for RVPm algorithms was 7.96% (95% CI 3.13–20.25), as compared with 45.11% (95% CI 26.64–76.38) of DDD pacing.
Conclusion
Algorithms for RVPm may be effective in reducing the risk of PerAF and CV hospitalization in patients requiring anti-bradycardia therapies, without an increased risk of adverse symptoms. These results are also consistent for studies enrolling a high percentage of AVB patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiology Division, Department of Biomedical, Metabolic and Neural Sciences, University of Modena and Reggio Emilia, Policlinico di Modena, Via del Pozzo 71, Modena 41121, Italy.
- Clinical and Experimental Medicine PhD Program, University of Modena and Reggio Emilia, Modena, Italy.
- Cardiology Unit, Department of Translational Medicine Sant'Anna University Hospital, University of Ferrara, Ferrara, Italy.
ORCID-Profile: Davide Antonio Mei, Jacopo Francesco Imberti, Marco Vitolo, Niccolò Bonini, Kevin Serafini, Marta Mantovani, Enrico Tartaglia, Marco Zuin, Matteo Bertini, Giuseppe Boriani
Zitieren
APA
Mei, D. A., Imberti, J. F., Vitolo, M., Bonini, N., Serafini, K., Mantovani, M., Tartaglia, E., Birtolo, C., Zuin, M., Bertini, M., & Boriani, G. (2024). Systematic review and meta-analysis on the impact on outcomes of device algorithms for minimizing right ventricular pacing. Europace, 26(8). https://doi.org/10.1093/europace/euae212
Vancouver
Mei DA, Imberti JF, Vitolo M, Bonini N, Serafini K, Mantovani M, et al. Systematic review and meta-analysis on the impact on outcomes of device algorithms for minimizing right ventricular pacing. Europace. 2024;26(8). doi:10.1093/europace/euae212
BibTeX
@article{Mei2024,
author = {Mei, Davide Antonio and Imberti, Jacopo Francesco and Vitolo, Marco and Bonini, Niccolò and Serafini, Kevin and Mantovani, Marta and Tartaglia, Enrico and Birtolo, Chiara and Zuin, Marco and Bertini, Matteo and Boriani, Giuseppe},
title = {Systematic review and meta-analysis on the impact on outcomes of device algorithms for minimizing right ventricular pacing},
journal = {Europace},
year = {2024},
volume = {26},
number = {8},
doi = {10.1093/europace/euae212}
}
EUVIMED