Inter‐ and Intra‐Rater Reliability of Pre‐Resection Ligament Tension Assessment Using a Digital Tensioner in Imageless Robotic‐Assisted Total Knee Replacement
Orthopaedic Surgery · 2026-05 · 18 (5) · 951-957
DOI: 10.1111/os.70299 ↗At a glance
Journal article in Orthopaedic Surgery, published online on 2026-03-25. Written by Xin Yang and 10 co-authors. Institution: Department of Orthopaedics and Traumatology, The Chinese University of Hong Kong, Hong Kong, SAR, China. Major MEDLINE topics: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Osteoarthritis, Knee, Ligaments, Articular. Cited 1 times so far (Crossref); 27 references. The full text is freely available (open access, CC BY-NC 4.0).
Abstract (original language)
Purpose
Ligament tension balance is essential for successful and long‐term implant survival of total knee arthroplasty (TKA). Conventional manual tensioning methods are subjective and vary with the surgeon experience. This study evaluated the inter‐ and intra‐rater reliability of pre‐resection ligament tension assessments using an imageless robotic‐assisted TKA system.
Methods
In this prospective study, 24 patients (30 knees) with end‐stage knee osteoarthritis underwent CORI‐assisted TKA from September 2024 to February 2025. Three specialist and three trainee surgeons each performed three repeated varus and valgus stress assessments across flexion and extension. Each medial and lateral gap measurement was recorded. Intraclass correlation coefficients (ICCs) were calculated with a two‐way mixed‐effects and two‐way random‐effects model.
Results
Both specialists and trainees demonstrated excellent intra‐ and inter‐ rater reliability (ICC > 0.90) for medial and lateral gaps in extension and medial gap in flexion. Flexion lateral reliability was good‐to‐excellent for trainees (ICC = 0.873, 95% CI: 0.784–0.933) and moderate‐to‐excellent for specialists (ICC = 0.838, 95% CI: 0.729–0.913).
Conclusions
Pre‐resection ligament tension assessment with an imageless robotic system yields high inter‐ and intra‐rater reliability, reducing variability linked to surgeon experience. Digital tensioners may standardize soft tissue balancing in TKA, potentially improving surgical outcomes and reproducibility. Evidence Level Level III, prospective quasi‐experimental study.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Orthopaedics and Traumatology, The Chinese University of Hong Kong, Hong Kong, SAR, China.
- Department of Orthopedic Surgery, The First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.
- Department of Orthopaedics and Traumatology, Prince of Wales Hospital, Hong Kong, SAR, China.
- Department of Orthopaedics and Traumatology, Alice Ho Miu Ling Nethersole Hospital, Hong Kong, SAR, China.
ORCID profiles: Rex Wang-Fung Mak, Mingde Cao, Patrick Shu Hang Yung, Michael Tim Yun Ong
Cite
APA
Yang, X., Xiao, L., Lei, P., Mak, R. W., Ng, J. P., Wong, C. K., Lam, G. Y., Choi, T. L., Cao, M., Yung, P. S. H., & Ong, M. T. Y. (2026). Inter‐ and Intra‐Rater Reliability of Pre‐Resection Ligament Tension Assessment Using a Digital Tensioner in Imageless Robotic‐Assisted Total Knee Replacement. Orthopaedic Surgery, 18(5), 951-957. https://doi.org/10.1111/os.70299
Vancouver
Yang X, Xiao L, Lei P, Mak RW, Ng JP, Wong CK, et al. Inter‐ and Intra‐Rater Reliability of Pre‐Resection Ligament Tension Assessment Using a Digital Tensioner in Imageless Robotic‐Assisted Total Knee Replacement. Orthopaedic Surgery. 2026;18(5):951-957. doi:10.1111/os.70299
BibTeX
@article{Yang2026,
author = {Yang, Xin and Xiao, Lingqing and Lei, Pengfei and Mak, Rex Wang-Fung and Ng, Jonathan Patrick and Wong, Cham-Kit and Lam, Gloria Yan-Ting and Choi, Tsz Lung and Cao, Mingde and Yung, Patrick Shu Hang and Ong, Michael Tim Yun},
title = {Inter‐ and Intra‐Rater Reliability of Pre‐Resection Ligament Tension Assessment Using a Digital Tensioner in Imageless Robotic‐Assisted Total Knee Replacement},
journal = {Orthopaedic Surgery},
year = {2026},
volume = {18},
number = {5},
pages = {951--957},
doi = {10.1111/os.70299}
}