Accuracy and precision of a sphero‐cylindrical over‐refraction app for smartphones

Rosa Maria Salmeron-Campillo iD, Gines Martinez-Ros, Jose Angel Diaz-Guirado, Carmen Travel-Alarcon, Mateusz Jaskulski, Norberto Lopez-Gil iD

Ophthalmic and Physiological Optics · 2025

DOI: 10.1111/opo.13560 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, 2025. Verfasst von Rosa Maria Salmeron-Campillo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Grupo de Ciencias de La Visión (CiViUM), Facultad de Óptica y Optometría, University of Murcia, Murcia, Spain. Hauptthemen laut MEDLINE: Smartphone, Mobile Applications, Refraction, Ocular, Refractive Errors. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the accuracy and precision of a web‐based smartphone app designed to measure sphero‐cylindrical over‐refraction.

Methods

A total of 307 healthy young subjects (22.5 ± 3.4 years) underwent clinical subjective refraction. In addition to each subject's refractive correction, spherical and cylindrical errors from −1.50 to 0.00 D were randomly induced so that their sum did not exceed −2.50 D, thereby generating a clinical over‐refraction. Seven different IOS and Android‐based smartphones were used. A new app showing tripole rescaling stimuli in blue or red light was used. Each subject used one smartphone to measure their induced over‐refraction subjectively, obtaining an app over‐refraction finding. Differences between the app and clinical over‐refraction were analysed across 675 values.

Results

The mean difference between the app and clinical over‐refraction in terms of power vector components ± limits of agreement was 0.17 ± 0.84 D for M ( p < 0.0001), 0.01 ± 0.42 D ( p = 0.28) for J 0 and −0.01 ± 0.32 D ( p = 0.33) for J 45 . The corresponding values for sphere and cylinder were 0.16 ± 0.80 D ( p < 0.0001) and 0.01 ± 0.85 D ( p = 0.40), respectively. The sphere and cylinder measurements obtained by the app differed from the expected values by no more than 0.50 D in 85% of cases.

Conclusions

The accuracy and precision of the over‐refraction obtained by the app was slightly lower than standard clinical subjective refraction in terms of inter‐examiner repeatability and comparable to measurements obtained using objective methods. The app shows the potential to perform sphero‐cylindrical over‐refractions reliably.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SmartphoneMobile ApplicationsRefraction, OcularRefractive ErrorsHumansMaleFemaleYoung AdultAdultReproducibility of ResultsHealthy Volunteers

Stichwörter der Autor:innen

mobile applicationover‐refractionrefractionsmartphone

Autor:innen und Institutionen

  • Grupo de Ciencias de La Visión (CiViUM), Facultad de Óptica y Optometría, University of Murcia, Murcia, Spain.
  • Visionapp Solutions S.L., Murcia, Spain.

ORCID-Profile: Rosa Maria Salmeron-Campillo, Norberto Lopez-Gil

Zitieren

APA

Salmeron-Campillo, R. M., Martinez-Ros, G., Diaz-Guirado, J. A., Travel-Alarcon, C., Jaskulski, M., & Lopez-Gil, N. (2025). Accuracy and precision of a sphero‐cylindrical over‐refraction app for smartphones. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.13560

Vancouver

Salmeron-Campillo RM, Martinez-Ros G, Diaz-Guirado JA, Travel-Alarcon C, Jaskulski M, Lopez-Gil N. Accuracy and precision of a sphero‐cylindrical over‐refraction app for smartphones. Ophthalmic and Physiological Optics. 2025. doi:10.1111/opo.13560

BibTeX
@article{SalmeronCampillo2025,
  author = {Salmeron-Campillo, Rosa Maria and Martinez-Ros, Gines and Diaz-Guirado, Jose Angel and Travel-Alarcon, Carmen and Jaskulski, Mateusz and Lopez-Gil, Norberto},
  title = {Accuracy and precision of a sphero‐cylindrical over‐refraction app for smartphones},
  journal = {Ophthalmic and Physiological Optics},
  year = {2025},
  doi = {10.1111/opo.13560}
}
Kostenlos starten →