Repeatability of lipid layer thickness using LipiView® following removal of contact lenses and its relationship to comfort
Ophthalmic and Physiological Optics · 2025
DOI: 10.1111/opo.13445 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, 2025. Verfasst von Mukesh Kumar und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Optometry and Vision Science, University of New South Wales, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Tears, Contact Lenses, Hydrophilic, Interferometry, Dry Eye Syndromes. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To assess the repeatability of lipid layer thickness (LLT) measurement using the LipiView® interferometer after daily disposable contact lens (CL) wear and correlation with ocular comfort in soft contact lens wearers.
Methods
A prospective study was conducted over two consecutive months, wherein CL wearers ( n = 20) wore either Somofilcon A or Verofilcon A daily disposable CLs in a crossover design, switching lenses after 1 month. The pre‐corneal tear film LLT was measured at the end of each month after CLs had been worn for at least 6 h. Using the LipiView® interferometer, three measurements of the LLT (average, maximum and minimum) were recorded. Subjects' symptoms were evaluated with the Contact Lens Dry Eye Questionnaire‐8 (CLDEQ‐8) and correlations sought between post‐lens pre‐corneal tear film LLT and symptoms.
Results
The average, maximum and minimum intraclass correlation coefficients (ICC) of LLT measurements at baseline were 0.57, 0.39 and 0.66, respectively, indicating poor (ICC < 0.4) to moderate (≥0.4,
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Optometry and Vision Science, University of New South Wales, Sydney, New South Wales, Australia.
- Brien Holden Vision Institute, University of New South Wales, Sydney, New South Wales, Australia.
- LV Prasad Eye Institute, Hyderabad, India.
ORCID-Profile: Mukesh Kumar
Zitieren
APA
Kumar, M., Masoudi, S., Vijay, A. K., Naduvilath, T. J., Dumpati, S., Raj, A., & Willcox, M. (2025). Repeatability of lipid layer thickness using LipiView® following removal of contact lenses and its relationship to comfort. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.13445
Vancouver
Kumar M, Masoudi S, Vijay AK, Naduvilath TJ, Dumpati S, Raj A, et al. Repeatability of lipid layer thickness using LipiView® following removal of contact lenses and its relationship to comfort. Ophthalmic and Physiological Optics. 2025. doi:10.1111/opo.13445
BibTeX
@article{Kumar2025,
author = {Kumar, Mukesh and Masoudi, Simin and Vijay, Ajay Kumar and Naduvilath, Thomas John and Dumpati, Srikanth and Raj, Ankit and Willcox, Mark},
title = {Repeatability of lipid layer thickness using LipiView® following removal of contact lenses and its relationship to comfort},
journal = {Ophthalmic and Physiological Optics},
year = {2025},
doi = {10.1111/opo.13445}
}
EUVIMED