EUVIMEDEUVIMED entdecken
Crossref

Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study

Fei Sun iD, Li-Bing Xu, Song-Xian Lai, Xin-Chang Li, Hai Xu, Zeng Lin

Journal of Prosthodontics · 2026

DOI: 10.1111/jopr.70062 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Prosthodontics, 2026. Verfasst von Fei Sun und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering and Automation, Northeastern University, Shenyang, Liaoning, China. Hauptthemen laut MEDLINE: Dental Implant-Abutment Design, Dental Abutments, Dental Implants. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Chinese Academy of Sciences-WEGO Research Development Plan, Fundamental Research Funds for the Central Universities of China, China Postdoctoral Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The purpose of this study is to analyze the impact of changes in the implant–abutment interface fit taper on loosening and fatigue performance. Materials and Methods Three abutment structures with varying tapers (11.5°, 12°, and 13°) were self‐designed and customized. Then, the impact of different taper fits between the abutment and implant on loosening and fatigue performance was investigated. The experiments measured screw preload, static and dynamic loosening torque, abutment settlement, pullout force, and the fatigue life of the implant system. The surface wear of the abutment and fatigue fracture surfaces were analyzed, and the experimental results were validated through finite element analysis (FEA).

Results

The abutment and implant have the same 11.5° matching taper increased preload, loosening torque, pullout force, and fatigue life. The fracture of the implant system occurs at the simulated bone plane, and the FEA of the stress concentration area aligns with the fatigue fracture results.

Conclusions

Adopting a consistent 11.5° taper design at the abutment–implant interface could enhance the anti‐loosening effect and extend the fatigue life, but a higher abutment pullout force may increase the difficulty of clinical restoration. Therefore, when the connection cone angle of the implant is 11.5°, using a 12° conical abutment may achieve optimal application results.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental Implant-Abutment DesignDental AbutmentsDental ImplantsTorqueFinite Element AnalysisSurface PropertiesDental Stress AnalysisDental Restoration FailureStress, MechanicalHumansMaterials Testing

Stichwörter der Autor:innen

FEAfatigue resistanceimplant‐abutment interfacescrew loosening

Substanzen

Dental Implants

Autor:innen und Institutionen

  • School of Mechanical Engineering and Automation, Northeastern University, Shenyang, Liaoning, China.
  • WEGO JERICOM Biomaterials Co., Ltd., Weihai, Shandong, China.

ORCID-Profile: Fei Sun

Förderung

  • Chinese Academy of Sciences-WEGO Research Development Plan
  • Fundamental Research Funds for the Central Universities of China
  • China Postdoctoral Science Foundation
  • Joint Fund of Natural Science Foundation of Liaoning Province
  • Northeastern University Postdoctoral Science Foundation
  • Fundamental Research Funds for the Central Universities

Zitieren

APA

Sun, F., Xu, L. B., Lai, S. X., Li, X. C., Xu, H., & Lin, Z. (2026). Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study. Journal of Prosthodontics. https://doi.org/10.1111/jopr.70062

Vancouver

Sun F, Xu LB, Lai SX, Li XC, Xu H, Lin Z. Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study. Journal of Prosthodontics. 2026. doi:10.1111/jopr.70062

BibTeX
@article{Sun2026,
  author = {Sun, Fei and Xu, Li-Bing and Lai, Song-Xian and Li, Xin-Chang and Xu, Hai and Lin, Zeng},
  title = {Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study},
  journal = {Journal of Prosthodontics},
  year = {2026},
  doi = {10.1111/jopr.70062}
}