Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study
Journal of Prosthodontics · 2026
DOI: 10.1111/jopr.70062 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Prosthodontics, 2026. Verfasst von Fei Sun und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering and Automation, Northeastern University, Shenyang, Liaoning, China. Hauptthemen laut MEDLINE: Dental Implant-Abutment Design, Dental Abutments, Dental Implants. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Chinese Academy of Sciences-WEGO Research Development Plan, Fundamental Research Funds for the Central Universities of China, China Postdoctoral Science Foundation.
Abstract (Originalsprache)
Purpose
The purpose of this study is to analyze the impact of changes in the implant–abutment interface fit taper on loosening and fatigue performance. Materials and Methods Three abutment structures with varying tapers (11.5°, 12°, and 13°) were self‐designed and customized. Then, the impact of different taper fits between the abutment and implant on loosening and fatigue performance was investigated. The experiments measured screw preload, static and dynamic loosening torque, abutment settlement, pullout force, and the fatigue life of the implant system. The surface wear of the abutment and fatigue fracture surfaces were analyzed, and the experimental results were validated through finite element analysis (FEA).
Results
The abutment and implant have the same 11.5° matching taper increased preload, loosening torque, pullout force, and fatigue life. The fracture of the implant system occurs at the simulated bone plane, and the FEA of the stress concentration area aligns with the fatigue fracture results.
Conclusions
Adopting a consistent 11.5° taper design at the abutment–implant interface could enhance the anti‐loosening effect and extend the fatigue life, but a higher abutment pullout force may increase the difficulty of clinical restoration. Therefore, when the connection cone angle of the implant is 11.5°, using a 12° conical abutment may achieve optimal application results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Mechanical Engineering and Automation, Northeastern University, Shenyang, Liaoning, China.
- WEGO JERICOM Biomaterials Co., Ltd., Weihai, Shandong, China.
ORCID-Profile: Fei Sun
Förderung
- Chinese Academy of Sciences-WEGO Research Development Plan
- Fundamental Research Funds for the Central Universities of China
- China Postdoctoral Science Foundation
- Joint Fund of Natural Science Foundation of Liaoning Province
- Northeastern University Postdoctoral Science Foundation
- Fundamental Research Funds for the Central Universities
Zitieren
APA
Sun, F., Xu, L. B., Lai, S. X., Li, X. C., Xu, H., & Lin, Z. (2026). Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study. Journal of Prosthodontics. https://doi.org/10.1111/jopr.70062
Vancouver
Sun F, Xu LB, Lai SX, Li XC, Xu H, Lin Z. Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study. Journal of Prosthodontics. 2026. doi:10.1111/jopr.70062
BibTeX
@article{Sun2026,
author = {Sun, Fei and Xu, Li-Bing and Lai, Song-Xian and Li, Xin-Chang and Xu, Hai and Lin, Zeng},
title = {Effect of taper fit at the dental implant–abutment interface on mechanical performance: In vitro study},
journal = {Journal of Prosthodontics},
year = {2026},
doi = {10.1111/jopr.70062}
}
EUVIMED