Vergleichsstudie

Trueness of 3D‐printed cobalt chromium versus titanium removable partial denture clasps

Mohammed E Elsayed, Dalia A El-Bayoumi, Ahmed M Hebeshi iD, Kawkb M El-Tamimi iD

Journal of Prosthodontics · 2024

DOI: 10.1111/jopr.13906 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Prosthodontics, 2024. Verfasst von Mohammed E Elsayed und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Dentistry, Suez Canal University, Ismailia, Egypt. Hauptthemen laut MEDLINE: Titanium, Chromium Alloys, Denture, Partial, Removable, Dental Clasps. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 42 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The purpose of this in vitro study was to evaluate the trueness of removable partial denture clasps fabricated with titanium (Ti) through the selective laser melting (SLM) technique compared to cobalt‐chromium alloys (CoCr). Materials and Methods A virtual Aker clasp was designed on a scanned tooth, and SLM printers were used to print 20 claps using cobalt chromium ( n = 10) and titanium alloy ( n = 10). The deviation between the printed clasps and reference design was measured using the surface matching software (Geomagic control x) at rest, retentive tip, reciprocal tip, retentive shoulder, and reciprocal shoulder. An Independent t ‐test was used to determine the influence of 3D‐printed material on the trueness ( a = 0.05).

Results

The gap distance in mm between the reference design and printed in titanium showed an average of 0.0001 ±0.0544, 0.0256 ±0.1309, 0.0230 ±0.1028, 0.0701 ±0.1234, and 0.0013 ±0.0735 mm in rest, reciprocal arm tip, retentive arm tip, retentive arm shoulder, and reciprocal arm shoulder, respectively. The gap distance in mm between the reference design and printed clasps in CoCr alloy showed an average of 0.0316 ±0.0692, 0.2783 ±0.1678, 0.1446 ±0.1528, 0.0315 ±0.0906, and 0.0419 ±0.1088 mm in rest, reciprocal arm tip, retentive arm tip, retentive arm shoulder, and reciprocal arm shoulder, respectively. The difference between titanium and CoCr alloys at each observation site was significant.

Conclusion

Clasps fabricated from titanium with SLM printing have the least deviation and better trueness compared to those fabricated from cobalt chromium.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TitaniumChromium AlloysDenture, Partial, RemovableDental ClaspsPrinting, Three-DimensionalDenture DesignDental MaterialsHumansSurface PropertiesMaterials TestingDenture RetentionLasersDental Alloys

Stichwörter der Autor:innen

claspdigital dentistrypartial denturetitanium printing

Substanzen

TitaniumChromium AlloysDental AlloysDental Materials

Autor:innen und Institutionen

  • Faculty of Dentistry, Suez Canal University, Ismailia, Egypt.
  • Faculty of Dentistry, Ain Shams University, Cairo, Egypt.

ORCID-Profile: Ahmed M Hebeshi, Kawkb M El-Tamimi

Zitieren

APA

Elsayed, M. E., El-Bayoumi, D. A., Hebeshi, A. M., & El-Tamimi, K. M. (2024). Trueness of 3D‐printed cobalt chromium versus titanium removable partial denture clasps. Journal of Prosthodontics. https://doi.org/10.1111/jopr.13906

Vancouver

Elsayed ME, El-Bayoumi DA, Hebeshi AM, El-Tamimi KM. Trueness of 3D‐printed cobalt chromium versus titanium removable partial denture clasps. Journal of Prosthodontics. 2024. doi:10.1111/jopr.13906

BibTeX
@article{Elsayed2024,
  author = {Elsayed, Mohammed E and El-Bayoumi, Dalia A and Hebeshi, Ahmed M and El-Tamimi, Kawkb M},
  title = {Trueness of 3D‐printed cobalt chromium versus titanium removable partial denture clasps},
  journal = {Journal of Prosthodontics},
  year = {2024},
  doi = {10.1111/jopr.13906}
}
Kostenlos starten →