EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Contact point marking with the OccluSense system—an in vitro study on reliability and validity

Stefanie Popa, M Oliver Ahlers iD

Journal of Oral Rehabilitation · 2024

DOI: 10.1111/joor.13774 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Oral Rehabilitation, 2024. Verfasst von Stefanie Popa und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthetic Dentistry, Center for Dental and Oral Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany. Hauptthemen laut MEDLINE: Dental Occlusion. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Occlusal contacts can be stained and thereby visualised employing occlusal indicators. Qualitative and quantitative indicators are differentiated. The hybrid system OccluSense supposedly combines analogue contact marking and digital registration of contacts. The reliability and validity of its contact marking have not yet been validated.

Objectives

The aim of this controlled randomised in‐vitro study was to examine the reliability and validity of analogue contact staining with OccluSense compared to Arti‐Fol and the reference standard Gnatho‐Film. Materials and Methods An experimental apparatus was designed to simulate static occlusion. The occlusal contacts were stained 50 times with Gnatho‐Film to determine the average number and location of contacts registered. For comparison, the contacts were pre‐stained with Gnatho‐Film and then registered with Arti‐Fol or OccluSense 50 times each. Every staining was statistically evaluated and compared to Gnatho‐Film alone, regarding the number and location of contacts. Based on the observed characteristic staining behaviour of OccluSense, a modified evaluation strategy was developed for its colour markings (OccluSense (mod.)).

Results

Both evaluations of OccluSense determine that the same number of contacts was registered in over 94% of all cases. Neither OccluSense nor Arti‐Fol stain the exact same number of contacts as Gnatho‐Film, but the modified evaluation OccluSense (mod.) did improve the results for validity.

Conclusion

When employing the modified evaluation strategy, OccluSense staining achieves similar reliability and validity as the reference standard. Pre‐colouring of occlusal contacts with Gnatho‐Film and the modified evaluation strategy, might become standard for analysing occlusal contacts stained by OccluSense in the future.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental OcclusionReproducibility of ResultsHumansJaw Relation RecordIn Vitro Techniques

Stichwörter der Autor:innen

apparatusdental instrumentsdental occlusiondigital technologyoral rehabilitationrestorative dentistry

Autor:innen und Institutionen

  • Department of Prosthetic Dentistry, Center for Dental and Oral Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • CMD-Centre Hamburg-Eppendorf, Hamburg, Germany.

ORCID-Profile: M Oliver Ahlers

Zitieren

APA

Popa, S., & Ahlers, M. O. (2024). Contact point marking with the OccluSense system—an in vitro study on reliability and validity. Journal of Oral Rehabilitation. https://doi.org/10.1111/joor.13774

Vancouver

Popa S, Ahlers MO. Contact point marking with the OccluSense system—an in vitro study on reliability and validity. Journal of Oral Rehabilitation. 2024. doi:10.1111/joor.13774

BibTeX
@article{Popa2024,
  author = {Popa, Stefanie and Ahlers, M Oliver},
  title = {Contact point marking with the OccluSense system—an in vitro study on reliability and validity},
  journal = {Journal of Oral Rehabilitation},
  year = {2024},
  doi = {10.1111/joor.13774}
}
Kostenlos starten →