A novel automated pipeline to assess MR spectroscopy quality control: Comparing current standards and manual assessment

Bodhi Beroukhim iD, Skyler McComas iD, Julie M Joyce, Luisa S Schuhmacher, Inga Koerte, Zhou Lan, Alexander Lin iD

Journal of Neuroimaging · 2025-01 · 35 (1)

DOI: 10.1111/jon.13246 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Neuroimaging, online erschienen am 05.11.2024. Verfasst von Bodhi Beroukhim und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center for Clinical Spectroscopy, Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Spectroscopy, Quality Control. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: NIH HHS, European Research Council, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Purpose The absence of a consensus data quality control (DQC) process inhibits the widespread adoption of MR spectroscopy. Poor DQC can lead to unreliable clinical diagnosis and irreproducible research conclusions. Currently, manual visual assessment or the standard quantitative metrics of signal‐to‐noise, linewidth, and model fit are used as classifiers, but these measures may not be sufficient. To supplement standard metrics, this paper proposes a novel automated DQC pipeline named Visual Evaluative Control Technology Of Resonance Spectroscopy (VECTORS).

Methods

Manual DQC ratings were conducted on 7180 spectra obtained from 110 young adults using short‐echo chemical shift imaging at 3 Tesla. Four reviewers conducted manual ratings on the presence of artifacts and location of metabolites. The ratings were labor intensive, taking over 180 hours. VECTORS was developed to quantify their DQC criteria, detecting artifacts that present as duplicate peaks, vertical shifts, and glutamine + glutamate and myoinositol peak shapes. Run on the same data using a standard laptop, VECTORS only took 2 hours.

Results

The manual ratings were not monotonic to the standard quantitative metrics. VECTORS correctly flagged spectra that the manual ratings missed. VECTORS accurately flagged an additional 126 poor DQ spectra that consensus cutoffs of the standard quantitative metrics deemed good DQ.

Conclusion

Standard quantitative metrics may not account for all DQC artifacts as they are not monotonic to the manual ratings. However, manual ratings are labor intensive, subjective, and irreproducible. VECTORS addresses these issues and should be used in conjunction with standard quantitative metrics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance SpectroscopyQuality ControlHumansAdultYoung AdultFemaleMaleArtifactsBrainReproducibility of Results

Stichwörter der Autor:innen

artifactsautomated pipelinedata quality controlmagnetic resonance spectroscopy

Autor:innen und Institutionen

  • Center for Clinical Spectroscopy, Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • cBRAIN, Department of Child and Adolescent Psychiatry, Psychosomatics, and Psychotherapy, Ludwig-Maximilians-Universität, Munich, Germany.
  • Center for Clinical Investigation, Department of Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.

ORCID-Profile: Bodhi Beroukhim, Skyler McComas, Alexander Lin

Förderung

  • NIH HHS
  • European Research Council
  • National Institutes of Health

Zitieren

APA

Beroukhim, B., McComas, S., Joyce, J. M., Schuhmacher, L. S., Koerte, I., Lan, Z., & Lin, A. (2025). A novel automated pipeline to assess MR spectroscopy quality control: Comparing current standards and manual assessment. Journal of Neuroimaging, 35(1). https://doi.org/10.1111/jon.13246

Vancouver

Beroukhim B, McComas S, Joyce JM, Schuhmacher LS, Koerte I, Lan Z, et al. A novel automated pipeline to assess MR spectroscopy quality control: Comparing current standards and manual assessment. Journal of Neuroimaging. 2025;35(1). doi:10.1111/jon.13246

BibTeX
@article{Beroukhim2025,
  author = {Beroukhim, Bodhi and McComas, Skyler and Joyce, Julie M and Schuhmacher, Luisa S and Koerte, Inga and Lan, Zhou and Lin, Alexander},
  title = {A novel automated pipeline to assess MR spectroscopy quality control: Comparing current standards and manual assessment},
  journal = {Journal of Neuroimaging},
  year = {2025},
  volume = {35},
  number = {1},
  doi = {10.1111/jon.13246}
}
Kostenlos starten →