The Post‐Marketing Real‐World Outpatient Clinical Validation of Electronic Blood Pressure Monitors
The Journal of Clinical Hypertension · 2026-02 · 28 (2)
DOI: 10.1111/jch.70222 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift The Journal of Clinical Hypertension, online erschienen am 23.02.2026. Verfasst von Zhiping Zhang und 12 weiteren Autor:innen. Beteiligte Einrichtung: Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, People's Republic of China. Hauptthemen laut MEDLINE: Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory, Hypertension. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Natural Science Foundation of China, Songjiang District Science and Technology Research (Medical and Health Field).
Abstract (Originalsprache)
This study evaluated the accuracy of a variety of home upper‐arm oscillometric blood pressure monitors (BPMs) from 448 participants against the FDA‐certified XYZ110 auscultatory device using simultaneous measurements. Descriptive statistics were used to summarize the key findings of the study. Device performance was evaluated referencing international standards (ISO 81060–2:2013), where a mean absolute difference of ≤5 mmHg was defined as “Accurate.” In addition, survey questionnaires completed by a subset of participants were analyzed to provide supplementary insights. A total of 448 consecutive outpatient clinical patients attending a routine clinical visit with their BPMs (male 215 and female 233) were eligible to participate in this study. The overall mean age of the participants was 62.35±12.59 years. Most of the BPMs included in this study were of the Omron brand (79.69%, n = 357), followed by Yuwell (5.8%, n = 26) and others. 76.32% of systolic blood pressure (SBP) measurements and 69.89% of diastolic blood pressure (DBP) measurements from BPMs exhibited differences of ≤5 mmHg compared to the reference calibrator. 70.89% ( n = 318) were accurate in measuring SBP, and 60.27% ( n = 270) were accurate in measuring DBP. Our study revealed that over 20% of BPMs exhibited discrepancies of more than 5 mmHg compared to a reference calibrator. Overall, the Omron U30, HEM‐7211, and U10 models demonstrated relatively higher accuracy in blood pressure measurements. Regular patient to patient validation of BPMs is crucial to ensure accurate measurements for daily use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, People's Republic of China.
- General Medicine Department of Fangsong Community Health Service Center of Songjiang District, Shanghai, People's Republic of China.
- General Medicine Department of Zhoujiadu Community Health Service Center of Pudong District, Shanghai, People's Republic of China.
- School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, People's Republic of China.
Förderung
- National Natural Science Foundation of China
- Songjiang District Science and Technology Research (Medical and Health Field)
Zitieren
APA
Zhang, Z., Fan, M., Zhao, Q., Liu, J., Su, L., Wang, J., Xiao, P., Fang, Z., Li, Y., Yue, Y., Jiang, R., Wang, W., & Zhang, Z. (2026). The Post‐Marketing Real‐World Outpatient Clinical Validation of Electronic Blood Pressure Monitors. The Journal of Clinical Hypertension, 28(2). https://doi.org/10.1111/jch.70222
Vancouver
Zhang Z, Fan M, Zhao Q, Liu J, Su L, Wang J, et al. The Post‐Marketing Real‐World Outpatient Clinical Validation of Electronic Blood Pressure Monitors. The Journal of Clinical Hypertension. 2026;28(2). doi:10.1111/jch.70222
BibTeX
@article{Zhang2026,
author = {Zhang, Zhiping and Fan, Muge and Zhao, Qingsong and Liu, Jiaqi and Su, Lingyun and Wang, Jinliang and Xiao, Peiyao and Fang, Ziyang and Li, Yikai and Yue, Yanbin and Jiang, Rong and Wang, Wei and Zhang, Zhi},
title = {The Post‐Marketing Real‐World Outpatient Clinical Validation of Electronic Blood Pressure Monitors},
journal = {The Journal of Clinical Hypertension},
year = {2026},
volume = {28},
number = {2},
doi = {10.1111/jch.70222}
}
EUVIMED