Left Atrial Voltage Attenuation Beyond Ablation Sites After Unipolar Pulsed‐Field Ablation for Atrial Fibrillation

Satoshi Osaki iD, Koichi Inoue iD, Tsuyoshi Mishima, Tatsuhisa Ozaki iD, Takuya Ohashi, Haruhiko Abe, Kuniyasu Ikeoka iD, Haruya Yamane iD, Masayuki Nakamura, Yasunori Ueda iD

Journal of Cardiovascular Electrophysiology · 2026

DOI: 10.1111/jce.70511 ↗

At a glance

Journal article in Journal of Cardiovascular Electrophysiology, 2026. Written by Satoshi Osaki and 9 co-authors. Institution: Department of Cardiology, National Hospital Organization Osaka National Hospital, Osaka, Japan. Major MEDLINE topics: Atrial Fibrillation, Catheter Ablation, Atrial Function, Left, Heart Atria. 8 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Introduction

Unlike most pulsed‐field ablation (PFA) systems, Affera (Medtronic) delivers unipolar energy between a lattice‐tip catheter and a cutaneous return patch, potentially generating a broader electric field. We hypothesized that this may attenuate voltage beyond ablated regions and tested this using spatially matched pre‐ and post‐ablation maps. Methods and Results We retrospectively studied five consecutive patients undergoing primary atrial fibrillation ablation with the Affera system. Pre‐ and immediate post‐ablation voltage maps were generated using CARTO 3 and OCTARAY (Biosense Webster) with identical settings during right atrial pacing. Because Affera lacks point‐level 3D coordinates for spatial matching, coordinates (x, y, z) and bipolar voltages were extracted from CARTO 3. Pre‐ and post‐ablation points—outside energy delivery sites, excluding the annulus and left atrial appendage—were spatially matched using a 2 mm mutual nearest‐neighbor criterion. This yielded 3772 pairs. Bipolar voltages were averaged by case. Mean bipolar voltage decreased in all cases from 1.01 ± 0.61 to 0.56 ± 0.29 mV, representing a 40.2% reduction (95% CI, 24.6%–55.8%; p = 0.044). In a distance‐stratified analysis, residual bipolar voltage (post/pre × 100) was lowest just beyond the ablated zone and increased progressively with distance from the nearest application tag center (16.5% at 10–15 mm to 57.8% at 25–30 mm; all p < 0.001), becoming numerically comparable to baseline beyond approximately 30 mm.

Conclusions

Affera‐based unipolar ablation produced acute bipolar voltage attenuation that extended into non‐ablated myocardium adjacent to the lesion set but diminished with distance and was largely confined within approximately 30 mm of the application sites.

Subjects

MeSH terms (MEDLINE)

Atrial FibrillationCatheter AblationAtrial Function, LeftHeart AtriaHumansRetrospective StudiesTreatment OutcomeFemaleAction PotentialsMaleAgedMiddle AgedElectrophysiologic Techniques, CardiacTime FactorsHeart RateIrreversible Electroporation Therapy

Author keywords

atrial fibrillationcatheter ablationleft atriumpulsed‐field ablationvoltage mapping

Authors and institutions

  • Department of Cardiology, National Hospital Organization Osaka National Hospital, Osaka, Japan.

ORCID profiles: Satoshi Osaki, Koichi Inoue, Tatsuhisa Ozaki, Kuniyasu Ikeoka, Haruya Yamane, Yasunori Ueda

Cite

APA

Osaki, S., Inoue, K., Mishima, T., Ozaki, T., Ohashi, T., Abe, H., Ikeoka, K., Yamane, H., Nakamura, M., & Ueda, Y. (2026). Left Atrial Voltage Attenuation Beyond Ablation Sites After Unipolar Pulsed‐Field Ablation for Atrial Fibrillation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70511

Vancouver

Osaki S, Inoue K, Mishima T, Ozaki T, Ohashi T, Abe H, et al. Left Atrial Voltage Attenuation Beyond Ablation Sites After Unipolar Pulsed‐Field Ablation for Atrial Fibrillation. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70511

BibTeX
@article{Osaki2026,
  author = {Osaki, Satoshi and Inoue, Koichi and Mishima, Tsuyoshi and Ozaki, Tatsuhisa and Ohashi, Takuya and Abe, Haruhiko and Ikeoka, Kuniyasu and Yamane, Haruya and Nakamura, Masayuki and Ueda, Yasunori},
  title = {Left Atrial Voltage Attenuation Beyond Ablation Sites After Unipolar Pulsed‐Field Ablation for Atrial Fibrillation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70511}
}
Start for free →