EUVIMEDEUVIMED entdecken
Crossref

Sheath Withdrawal May Represent a Mechanically Vulnerable Phase for Septal Perforation During Left Bundle Branch Area Pacing

Masayuki Ohta iD, Yuki Ishidoya iD, Ryosuke Ito, Daisuke Sunaga, Naohiro Funayama iD, Katsuhiko Ohori, Yusuke Kashiwagi, Shunsuke Ohori

Journal of Cardiovascular Electrophysiology · 2026-06 · 37 (6) · 1286-1294

DOI: 10.1111/jce.70346 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 19.04.2026. Verfasst von Masayuki Ohta und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Hokkaido Cardiovascular Hospital, Sapporo, Japan. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Models, Cardiovascular, Pacemaker, Artificial, Bundle of His. Die Arbeit nennt 3 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Although interventricular septal perforation is a recognized complication of left bundle branch area pacing (LBBAP), its underlying mechanisms remain unclear. Previous studies have focused primarily on lead deployment‐related factors, whereas mechanical events occurring after apparently successful lead fixation have received limited attention.

Methods

In this exploratory mechanistic study, we investigated axial force transmission during delivery sheath withdrawal using a simplified benchtop model motivated by a representative clinical observation in which overt septal perforation became apparent immediately after sheath slitting. Axial force transmitted along the lead‐sheath system was quantified under two geometric configurations simulating a constrained (5 cm) and less‐constrained (10 cm) atrial space. Sheath withdrawal was performed in 1‐cm increments under three initial orientations (neutral, clockwise, and counterclockwise). Peak axial force values were analyzed using two‐way analysis of variance.

Results

In the benchtop model, the constrained configuration generated significantly higher peak axial forces than the less‐constrained configuration across all sheath orientations ( p < 0.001). Initial sheath orientation significantly modified force transmission, with a counterclockwise orientation producing the highest peak axial force under constrained conditions ( p < 0.001). Peak axial force consistently occurred during the early phase of sheath withdrawal, corresponding to a lead tip‐to‐sheath tip distance of approximately 3.5–4.5 cm. These mechanical findings were concordant with fluoroscopic observations from the index clinical case, in which distinct lead‐tip excursion and overt septal perforation were recognized during early sheath withdrawal. However, subtle pre‐existing protrusion of the lead helix before sheath withdrawal could not be excluded.

Conclusions

Sheath withdrawal during LBBAP may represent a mechanically vulnerable phase during which geometric constraint and sheath orientation can synergistically amplify axial force transmission to the lead tip. Recognition of this vulnerable phase, together with attention to atrial geometry and sheath orientation during early withdrawal, may help reduce the risk of lead destabilization or overt interventricular septal perforation during LBBAP.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Pacing, ArtificialModels, CardiovascularPacemaker, ArtificialBundle of HisVentricular Septal RuptureHumansStress, MechanicalRisk FactorsBiomechanical PhenomenaTime FactorsComputer Simulation

Stichwörter der Autor:innen

axial forcebenchtop modelcardiac pacing complicationsdelivery sheathleft bundle branch area pacingseptal perforationsheath withdrawal

Autor:innen und Institutionen

  • Department of Cardiology, Hokkaido Cardiovascular Hospital, Sapporo, Japan.

ORCID-Profile: Masayuki Ohta, Yuki Ishidoya, Naohiro Funayama

Zitieren

APA

Ohta, M., Ishidoya, Y., Ito, R., Sunaga, D., Funayama, N., Ohori, K., Kashiwagi, Y., & Ohori, S. (2026). Sheath Withdrawal May Represent a Mechanically Vulnerable Phase for Septal Perforation During Left Bundle Branch Area Pacing. Journal of Cardiovascular Electrophysiology, 37(6), 1286-1294. https://doi.org/10.1111/jce.70346

Vancouver

Ohta M, Ishidoya Y, Ito R, Sunaga D, Funayama N, Ohori K, et al. Sheath Withdrawal May Represent a Mechanically Vulnerable Phase for Septal Perforation During Left Bundle Branch Area Pacing. Journal of Cardiovascular Electrophysiology. 2026;37(6):1286-1294. doi:10.1111/jce.70346

BibTeX
@article{Ohta2026,
  author = {Ohta, Masayuki and Ishidoya, Yuki and Ito, Ryosuke and Sunaga, Daisuke and Funayama, Naohiro and Ohori, Katsuhiko and Kashiwagi, Yusuke and Ohori, Shunsuke},
  title = {Sheath Withdrawal May Represent a Mechanically Vulnerable Phase for Septal Perforation During Left Bundle Branch Area Pacing},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  volume = {37},
  number = {6},
  pages = {1286--1294},
  doi = {10.1111/jce.70346}
}
Kostenlos starten →