Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only

Caroline Wiederkehr iD, Noah Kuster, Michelle Bebié, Jan Steffel, Andreas Haeberlin, Alexander Breitenstein, Daniel Hofer iD

Journal of Cardiovascular Electrophysiology · 2026-06 · 37 (6) · 1241-1248

DOI: 10.1111/jce.70342 ↗

At a glance

Journal article in Journal of Cardiovascular Electrophysiology, published online on 2026-04-11. Written by Caroline Wiederkehr and 6 co-authors. Institution: Department of Cardiology, Triemli Hospital Zurich, Zurich, Switzerland. Major MEDLINE topics: Device Lead Extraction, Pacemaker, Artificial, Defibrillators, Implantable, Device Removal. 32 references. The full text is freely available (open access, CC BY-NC 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background and Aims Transvenous lead extraction (TLE) is increasingly performed for malfunction or infection of cardiac implantable electronic devices (CIED). Active mechanical rotational sheaths facilitate extraction and are available in short and long sizes, but it is unclear how many patients short active mechanical extraction sheaths (SAMES) are sufficient for TLE, and in which patients they may facilitate TLE. This study aimed to evaluate the incidence and predictors of successful lead extraction using only SAMES.

Methods

We retrospectively analyzed patients who underwent TLE at the University Hospital Zurich from 2013 until 2023 regarding the use of short or long active mechanical extraction sheaths, as well as clinical and radiological success and complications.

Results

A total of 568 TLEs were performed from January 2013 until December 2023. In 55 cases (9.7%), TLE was successful using only SAMES. Independent predictors for successful lead extraction using only SAMES in multivariate analysis were shorter lead dwell time (OR: 0.98, 95% CI: 0.97–0.99, p < 0.001) and use of anticoagulation (OR: 0.44, 95% CI: 0.19–1.00, p = 0.05). Complications occurred at a lower rate when only SAMES were used (All complications 12.7% vs. 24.2%, minor complications 9.1% vs. 18.7%, perioperative death 0% vs. 1.7%).

Conclusion

In 10% of patients undergoing TLE, SAMES are sufficient for successful lead extraction. SAMES may be a sensible choice as the primary extraction tool in patients with chronic anticoagulation and shorter lead dwell time, as these factors predict a higher likelihood of success in TLE with SAMES only.

Subjects

MeSH terms (MEDLINE)

Device Lead ExtractionPacemaker, ArtificialDefibrillators, ImplantableDevice RemovalProsthesis-Related InfectionsCardiac CathetersHumansRetrospective StudiesIncidenceRisk FactorsTreatment OutcomeTime FactorsFemaleMaleAgedEquipment DesignSwitzerlandMiddle AgedAged, 80 and over

Author keywords

active mechanical sheathscomplicationslead extractionoutcome

Authors and institutions

  • Department of Cardiology, Triemli Hospital Zurich, Zurich, Switzerland.
  • Department of Cardiology, University Hospital of Zurich, Division of Electrophysiology, Zurich, Switzerland.
  • Hirslanden Heart Clinic, Zurich, Switzerland.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

ORCID profiles: Caroline Wiederkehr, Daniel Hofer

Cite

APA

Wiederkehr, C., Kuster, N., Bebié, M., Steffel, J., Haeberlin, A., Breitenstein, A., & Hofer, D. (2026). Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only. Journal of Cardiovascular Electrophysiology, 37(6), 1241-1248. https://doi.org/10.1111/jce.70342

Vancouver

Wiederkehr C, Kuster N, Bebié M, Steffel J, Haeberlin A, Breitenstein A, et al. Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only. Journal of Cardiovascular Electrophysiology. 2026;37(6):1241-1248. doi:10.1111/jce.70342

BibTeX
@article{Wiederkehr2026,
  author = {Wiederkehr, Caroline and Kuster, Noah and Bebié, Michelle and Steffel, Jan and Haeberlin, Andreas and Breitenstein, Alexander and Hofer, Daniel},
  title = {Incidence and Predictors of Successful Lead Extraction With Short Active Mechanical Extraction Sheaths Only},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  volume = {37},
  number = {6},
  pages = {1241--1248},
  doi = {10.1111/jce.70342}
}
Start for free →