Distance Between the Electrode Tip and the Left Ventricular Conduction System During Left Bundle Branch Area Pacing
Journal of Cardiovascular Electrophysiology · 2026-04 · 37 (4) · 827-836
DOI: 10.1111/jce.70302 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 26.02.2026. Verfasst von Fan Zhang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Chest Hospital, Tianjin University, Tianjin, China. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Bundle of His, Heart Conduction System, Electrophysiologic Techniques, Cardiac. Die Arbeit nennt 28 Literaturangaben. Gefördert von: Natural Science for Youth Foundation, Tianjin Health Science and Technology Project, Tianjin Key Medical Discipline Construction Project.
Abstract (Originalsprache)
Introduction
Left bundle branch area pacing (LBBAP) represents an essential physiological pacing technique, and conduction system (CS) capture can be achieved even in the absence of detectable CS potential. We aimed to measure the distance between the electrode tip and the left ventricular conduction system (LVCS) during LBBAP. Methods and Results Sixteen canines are evenly divided into two groups: (1) conduction system pacing (CSP) group, including left bundle branch pacing (LBBP) and left fascicular pacing (LFP); (2) left ventricular septal pacing (LVSP) group. A three‐dimensional electroanatomical mapping (EAM) system used to visualize the LVCS and guide electrode implantation. Pathological specimens are used to measure the distances between the electrode tip and LVCS, as well as the depth and angle of electrode implantation. Group 1 showed electrode tip‐to‐LVCS distance ≤ 2 mm versus > 2 mm in Group 2. Significant differences existed in pacing and pathological parameters between groups: stimulus to left ventricular activation time (Stim‐LVAT) (48.75 ± 2.38 ms vs. 59.75 ± 2.05 ms, p < 0.01), Stim‐Retro‐His interval (21.00 ± 2.27 ms vs. 44.50 ± 2.83 ms, p < 0.01), QRS duration (89.50 ± 4.34 ms vs. 109.88 ± 5.96 ms, p < 0.01), the depth of electrode implantation (10.38 ± 0.53 mm vs. 9.35 ± 0.76 mm, p < 0.01), the angle of electrode implantation (76.88° ± 11.32° vs. 55.63° ± 12.66°, p < 0.01), and the distance between the electrode tip and the LVCS (0.69 ± 0.66 mm vs. 3.06 ± 0.28 mm, p < 0.01).
Conclusion
When the distance between the electrode tip and LVCS is ≤ 2 mm, the CS could be captured at outputs of ≤ 1.5 V/0.5 ms.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Chest Hospital, Tianjin University, Tianjin, China.
ORCID-Profile: Wei Ma
Förderung
- Natural Science for Youth Foundation
- Tianjin Health Science and Technology Project
- Tianjin Key Medical Discipline Construction Project
Zitieren
APA
Zhang, F., Ma, W., Fan, S., He, L., Yao, Y., Zhang, J., Zhang, X., Zhang, E., Lu, F., & Xu, J. (2026). Distance Between the Electrode Tip and the Left Ventricular Conduction System During Left Bundle Branch Area Pacing. Journal of Cardiovascular Electrophysiology, 37(4), 827-836. https://doi.org/10.1111/jce.70302
Vancouver
Zhang F, Ma W, Fan S, He L, Yao Y, Zhang J, et al. Distance Between the Electrode Tip and the Left Ventricular Conduction System During Left Bundle Branch Area Pacing. Journal of Cardiovascular Electrophysiology. 2026;37(4):827-836. doi:10.1111/jce.70302
BibTeX
@article{Zhang2026,
author = {Zhang, Fan and Ma, Wei and Fan, Shaobo and He, Le and Yao, Yanhong and Zhang, Jie and Zhang, Xin and Zhang, Enyuan and Lu, Fengmin and Xu, Jing},
title = {Distance Between the Electrode Tip and the Left Ventricular Conduction System During Left Bundle Branch Area Pacing},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
volume = {37},
number = {4},
pages = {827--836},
doi = {10.1111/jce.70302}
}
EUVIMED