Pacing system analyzers to guide conduction system pacing implantation procedures: A comparison study of intracardiac and surface signals with an electrophysiology recording system
Heart Rhythm · 2025-07 · 22 (7) · e113-e121
DOI: 10.1016/j.hrthm.2024.12.030 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Heart Rhythm, 2025-07. Verfasst von Jesús Jiménez-López und 18 weiteren Autor:innen. Beteiligte Einrichtung: Electrophysiology Unit, Cardiovascular Division, Hospital del Mar, Barcelona, Spain. Electronic address: jejilo78@gmail.com. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Electrophysiologic Techniques, Cardiac, Heart Conduction System, Electrocardiography. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 15 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Left bundle branch area pacing (LBBAP) needs confirmation of left conduction system capture assessed by testing of different electrical parameters. Guidelines recommend the use of an electrophysiology recording system (EP-RS) to guide conduction system pacing procedures. However, some experienced centers perform LBBAP procedures without an EP-RS. OBJECTIVE: We aimed to assess whether LBBAP criteria can be measured by the signals provided by the pacing system analyzers (PSAs) as a surrogate for the EP-RS to simplify and universalize the technique. METHODS: This was an observational, prospective, multicenter study assessing the current LBBAP criteria using the PSA compared with the EP-RS during conduction system pacing procedures. RESULTS: A consecutive 108 patients were included. Baseline QRS duration was 130 ± 29 ms in the EP-RS vs 128 ± 29 ms in the PSA (P = .7). An initial W morphology in V1 was noted in 88% of patients with the EP-RS vs 86% of patients with the PSA (P = .7) during unipolar pacing. The QRS duration (122 ± 17 ms vs 123 ± 19 ms; P = .7), the R-wave peak time interval in V6 (80 ± 13 ms vs 79 ± 14 ms (P = .9), and the V6-V1 interpeak interval (39 ± 16 ms vs 38 ± 17 ms (P = .7) were superimposable in the EP-RS compared with the PSA. Pearson coefficients for the last 2 criteria were 0.85 (P < .0001) and 0.94 (P < .0001), respectively. According to the current criteria, 91.5% of patients received a successful LBBAP implant using the EP-RS. Based on the PSA measurements, 96.6% of these patients met LBBAP criteria. CONCLUSION: Criteria for LBBAP can be assessed by PSAs with high accuracy. These results provide the basis for the usefulness of the PSA to guide LBBAP procedures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Electrophysiology Unit, Cardiovascular Division, Hospital del Mar, Barcelona, Spain. Electronic address: jejilo78@gmail.com.
- Electrophysiology Unit, Cardiovascular Division, Hospital del Mar, Barcelona, Spain; Department of Medicine, Universitat Autònoma de Barcelona, Bellaterra, Spain.
- Electrophysiology Unit, Cardiovascular Division, Hospital del Mar, Barcelona, Spain.
- Unitat d'Arritmies, Cardiology Department. University Hospital Vall d'Hebrón and Vall d'Hebron Research Institute, Autonomous University of Barcelona, CIBER-CV, Barcelona, Spain.
- Arrhythmia Unit, Cardiovascular Division, University Hospital Joan XXIII, Tarragona, Spain.
- Arrhythmia Unit, Hospital Virgen de las Nieves, Granada, Spain.
Zitieren
APA
Jiménez-López, J., González-Matos, C. E., Rodriguez-Queraltó, O., Francisco-Pascual, J., Lal-Trehan Estrada, N., Martín-Sánchez, G., Cabrera-Gómez, S., Lerma, M. M., Casteigt, B., Mercè-Klein, J., Rodríguez-García, M. A., Benito-Martín, E., Conejos-Montenegro, J., Rodríguez Garrido, J. L., Teja, J. E., Guillen, P. C., Silva, A. T., Reitmaier, C. L., & Gras, E. V. (2025). Pacing system analyzers to guide conduction system pacing implantation procedures: A comparison study of intracardiac and surface signals with an electrophysiology recording system. Heart Rhythm, 22(7), e113-e121. https://doi.org/10.1016/j.hrthm.2024.12.030
Vancouver
Jiménez-López J, González-Matos CE, Rodriguez-Queraltó O, Francisco-Pascual J, Lal-Trehan Estrada N, Martín-Sánchez G, et al. Pacing system analyzers to guide conduction system pacing implantation procedures: A comparison study of intracardiac and surface signals with an electrophysiology recording system. Heart Rhythm. 2025;22(7):e113-e121. doi:10.1016/j.hrthm.2024.12.030
BibTeX
@article{JimnezLpez2025,
author = {Jiménez-López, Jesús and González-Matos, Carlos E and Rodriguez-Queraltó, Oriol and Francisco-Pascual, Jaume and Lal-Trehan Estrada, Nisha and Martín-Sánchez, Gabriel and Cabrera-Gómez, Sandra and Lerma, Manuel Molina and Casteigt, Benjamín and Mercè-Klein, Jordi and Rodríguez-García, Marcos A and Benito-Martín, Eva and Conejos-Montenegro, Javier and Rodríguez Garrido, Jorge Luis and Teja, Juliana Elices and Guillen, Pau Cosialls and Silva, Aitor Toledo and Reitmaier, Carolina López and Gras, Ermengol Vallès},
title = {Pacing system analyzers to guide conduction system pacing implantation procedures: A comparison study of intracardiac and surface signals with an electrophysiology recording system},
journal = {Heart Rhythm},
year = {2025},
volume = {22},
number = {7},
pages = {e113--e121},
doi = {10.1016/j.hrthm.2024.12.030}
}
EUVIMED