Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Shayekh Abedin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Westmead Hospital, Sydney, Australia. Hauptthemen laut MEDLINE: Esophagus, Catheter Ablation, Pulmonary Veins, Burns. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: P.C.Q. was supported by a National Health and Medical Research Council (NHMRC) investigator grant (2018376), the Heart Foundation Future Leader Fellowship with….
Abstract (Originalsprache)
Introduction
Atrio‐esophageal fistula is a rare but devastating complication from thermal esophageal injury during pulmonary vein isolation (PVI). Modifying ablation energy delivery and placement of esophageal temperature probes (ETPs) have been employed to mitigate the risk of thermal esophageal injury. Our study sought to compare the effects of return electrode location, insulation of the ETP, and use of high‐power short‐duration (HPSD) ablation on esophageal temperature, ETP temperature measurement, and thermal lesion dimensions.
Methods
A myocardial phantom gel model using thermochromic material was used to approximate the atrio‐esophageal environment. Irrigated ablations were performed at 30 W/20 s (low‐power long duration [LPLD]) and 50 W/5 s (HPSD) with variation of return electrode position and insulation of the ETP. Temperatures were continuously logged, and digital photography with in‐house software was performed to examine lesion dimensions.
Results
Distant placement of a return electrode compared to proximal placement decreased peak esophageal temperature ( T Peak ) across LPLD (41.73°C ± 0.12°C vs. 39.23°C ± 0.15°C; p < 0.0001) and HPSD ablations (39.17°C ± 0.06°C vs. 38.13°C ± 0.06°C; p < 0.0001). Insulation of the ETP reduced T Peak across LPLD (39.13°C ± 0.25°C vs. 41.73°C ± 0.12°C; p < 0.0001) and HPSD ablation (39.17°C ± 0.06°C vs. 38.27°C ± 0.06°C; p < 0.0001). HPSD ablation produced lower mean T Peak than LPLD regardless of the location of the return electrode location or insulation of the ETP.
Conclusion
HPSD settings, insulation of the ETP, and a distant return electrode placement minimized esophageal heating. These strategies may reduce thermal injury to the esophagus during radiofrequency ablation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Westmead Hospital, Sydney, Australia.
- Westmead Applied Research Centre, Faculty of Medicine and Health, University of Sydney, Sydney, Australia.
ORCID-Profile: Xingzhou Liu
Förderung
- P.C.Q. was supported by a National Health and Medical Research Council (NHMRC) investigator grant (2018376), the Heart Foundation Future Leader Fellowship with…
Zitieren
APA
Abedin, S., Liu, X., Barry, M. A., Thomas, S. P., & Qian, P. C. (2026). Minimizing Esophageal Heating During Radiofrequency Catheter Ablation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70158
Vancouver
Abedin S, Liu X, Barry MA, Thomas SP, Qian PC. Minimizing Esophageal Heating During Radiofrequency Catheter Ablation. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70158
BibTeX
@article{Abedin2026,
author = {Abedin, Shayekh and Liu, Xingzhou and Barry, Michael A and Thomas, Stuart P and Qian, Pierre C},
title = {Minimizing Esophageal Heating During Radiofrequency Catheter Ablation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
doi = {10.1111/jce.70158}
}
EUVIMED