Fallbericht in Journal of Cardiovascular Electrophysiology, 2025. Written by Malte Kranert and 4 co-authors. Institution: Department of Cardiology and Angiology, University Hospital of Tuebingen, Tuebingen, Germany. Major MEDLINE topics: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation. 10 references. The full text is freely available (open access, CC BY 4.0). Funded by: The authors received no specific funding for this work..
Abstract (original language)
Introduction
Pulmonary vein isolation (PVI) in atrial fibrillation (AF) is the current cornerstone for AF ablation. Besides thermal ablation techniques, nonthermal pulsed‐field ablation (PFA) is rapidly transforming the way PVI is achieved. As every new technology has its own learning curve, we would like to report the lessons we learned by treating a 63‐year‐old woman admitted for paroxysmal AF ablation, with symptomatic palpitations. Methods and Results We initially performed PVI with a pulsed‐electrical field variable‐loop circular catheter with the CARTO 3 mapping system. When the patient experienced recurrence of AF even after 3 months, the scheduled reablation procedure showed a reconnection of both posterior carinae. When revisiting the initial map, both recurrences occurred at the points where tissue proximity indication (TPI) during ablation was likely not sufficient, leading to reversible electroporation and subsequent reconnection of the veins. Those remaining areas were again ablated using PFA energy with a linear ablation catheter.
Conclusion
TPI‐based contact filtering for transmural lesion success and durability seems to be a relevant tool to ensure irreversible electroporation and is likely of prognostic relevance.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiology and Angiology, University Hospital of Tuebingen, Tuebingen, Germany.
- Department of Pediatric Cardiology, Pulmonology and Intensive Care Medicine, University Children's Hospital, Tuebingen, Germany.
ORCID profiles: Malte Kranert, David Heinzmann
Funding
- The authors received no specific funding for this work.
Cite
APA
Kranert, M., Scheckenbach, C., Harm, T., Gawaz, M., & Heinzmann, D. (2025). Pulsed‐Electrical Field Ablation Completes Pulsed‐Field Ablation—Lessons Learned on Tissue Proximity Indication in Pulmonary Vein Isolation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70026
Vancouver
Kranert M, Scheckenbach C, Harm T, Gawaz M, Heinzmann D. Pulsed‐Electrical Field Ablation Completes Pulsed‐Field Ablation—Lessons Learned on Tissue Proximity Indication in Pulmonary Vein Isolation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70026
BibTeX
@article{Kranert2025,
author = {Kranert, Malte and Scheckenbach, Christian and Harm, Tobias and Gawaz, Meinrad and Heinzmann, David},
title = {Pulsed‐Electrical Field Ablation Completes Pulsed‐Field Ablation—Lessons Learned on Tissue Proximity Indication in Pulmonary Vein Isolation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.70026}
}
EUVIMED