Outcomes of Exclusive 50 W High Power, 5 s Ablation Close to the Esophagus in Pulmonary Vein Isolation
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.16678 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Yoshikazu Sato und 9 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Center, Tsuchiura Kyodo Hospital, Tsuchiura, Ibaraki, Japan. Hauptthemen laut MEDLINE: Pulmonary Veins, Catheter Ablation, Esophagus, Atrial Fibrillation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
In pulmonary vein isolation (PVI) employing a high‐power and short‐duration (HP‐SD) strategy, no consensus exists regarding the optimal settings for radiofrequency application (RFA) to the section of left atrium adjacent to the esophagus (esophageal area, EA).
Objective
To evaluate (1) the efficacy and safety of HP‐SD RFA with 50 W for 5 s on the EA, and (2) the predictive factors of esophageal thermal injury (ETI).
Methods
We performed esophagogastroduodenoscopy on 85 patients who had undergone PVI with 50 W RFA. RFA duration was limited to 5 s on the EA.
Results
Of the 85 patients, 16 (18.8%) developed asymptomatic thermal injury (1 mucosal injury, namely erosion and 15 gastric hypomotility). All of those were asymptomatic and transient, and any symptomatic or severe, esophageal or gastric complications was not observed. On the EA, conduction block was achieved after the first encircling in 82 (96.5%) patients. The number of RFA on the EA and the proportion of patients whose EA was located at the posterior wall of the right inferior PV (right‐sided EA) were higher in the patients with ETI than without (9.0 [7.0−10.5] vs. 7.0 [5.0−9.0] p = 0.021, and 18.8% vs. 2.9%, p = 0.044, respectively). In multivariate analysis, the number of RFA on the EA (hazard ratio 1.32, p = 0.019) and RFA on the right‐sided EA (hazard ratio, 13.50, p = 0.013) were identified as the predictors of ETI.
Conclusion
RFA of 50 W for 5 s on the EA can be performed with a low incidence of esophageal mucosal injury while maintaining efficacy. RFA on the right‐sided EA and the high number of RFA on the EA increase the risk of ETI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Center, Tsuchiura Kyodo Hospital, Tsuchiura, Ibaraki, Japan.
- Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan.
ORCID-Profile: Yoshikazu Sato, Shigeki Kusa, Satoshi Hara, Tadanori Nakata, Tetsuo Sasano, Hitoshi Hachiya
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Sato, Y., Kusa, S., Miwa, N., Hara, S., Nakata, T., Hirano, H., Ishizawa, T., Tahara, T., Sasano, T., & Hachiya, H. (2025). Outcomes of Exclusive 50 W High Power, 5 s Ablation Close to the Esophagus in Pulmonary Vein Isolation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16678
Vancouver
Sato Y, Kusa S, Miwa N, Hara S, Nakata T, Hirano H, et al. Outcomes of Exclusive 50 W High Power, 5 s Ablation Close to the Esophagus in Pulmonary Vein Isolation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.16678
BibTeX
@article{Sato2025,
author = {Sato, Yoshikazu and Kusa, Shigeki and Miwa, Naoyuki and Hara, Satoshi and Nakata, Tadanori and Hirano, Hidenori and Ishizawa, Taiki and Tahara, Tomohiro and Sasano, Tetsuo and Hachiya, Hitoshi},
title = {Outcomes of Exclusive 50 W High Power, 5 s Ablation Close to the Esophagus in Pulmonary Vein Isolation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.16678}
}
EUVIMED