Comparison of efficiency of PFA catheter designs by computer modeling

Andres Belalcazar iD, E Kevin Heist

Journal of Cardiovascular Electrophysiology · 2024-12 · 35 (12) · 2382-2393

DOI: 10.1111/jce.16459 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 08.10.2024. Verfasst von Andres Belalcazar und 1 weiteren Autor:innen. Beteiligte Einrichtung: Biophysicist, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Cardiac Catheters, Equipment Design, Catheter Ablation, Models, Cardiovascular. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: None.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Various catheter designs are appearing for Pulsed Field Ablation (PFA). It is unclear if they differ in terms of safety and efficiency. PFA studies have reported hemolysis, kidney injury, high troponin, among other side effects.

Methods

Using a CT‐derived computer model, we compared catheter designs using two metrics: (1) efficiency: power delivered to an atrial wall target, expressed as a percent of total generator power; and (2) safety: electric current to achieve 90% transmurality (since more energy causes more collateral effects), as well as the corresponding electrode current density (ECD), a heat and bubble metric. The following catheter designs were compared: penta‐spline basket, Nitinol spheres (focal 9 mm and large 1‐shot), circular, balloon, and flex‐circuit. Target was a 6 × 47 mm circumferential segment of atrial wall at LPV antrum. Transmurality was defined as percent of target having >600 volts per centimeter (V/cm) electric field needed for electroporation.

Results

Efficiency was 0.9, 1.4, 2.7, 5.9, 10, and 12% for the large 1‐shot and 9 mm Nitinol spheres, penta‐spline, circular, flex spline, and balloon catheters, respectively. Regarding safety, currents for 90% transmurality were 70, 39,36,12.5, 5.3, and 4 Amps for the same respective catheters, with less being safer. ECD was 124, 25, 74, 83, 41, and 31 A/cm 2 , respectively.

Conclusion

Computer models demonstrated a remarkable range in efficiency among catheters studied. Those having less atrial blood exposure had the highest efficiencies, with factors of up to 13X more efficiency compared to exposed ones. Higher efficiency designs have less collateral current and are safer. Confirmatory in‐vivo studies are required.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac CathetersEquipment DesignCatheter AblationModels, CardiovascularHumansComputer SimulationHeart AtriaAlloysTomography, X-Ray Computed

Stichwörter der Autor:innen

atrial Fibrillationelectroporationpulmonary vein isolationpulsed field ablation

Substanzen

Alloys

Autor:innen und Institutionen

  • Biophysicist, Minneapolis, Minnesota, USA.
  • Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, Massachusetts, USA.

ORCID-Profile: Andres Belalcazar

Förderung

  • None

Zitieren

APA

Belalcazar, A., & Heist, E. K. (2024). Comparison of efficiency of PFA catheter designs by computer modeling. Journal of Cardiovascular Electrophysiology, 35(12), 2382-2393. https://doi.org/10.1111/jce.16459

Vancouver

Belalcazar A, Heist EK. Comparison of efficiency of PFA catheter designs by computer modeling. Journal of Cardiovascular Electrophysiology. 2024;35(12):2382-2393. doi:10.1111/jce.16459

BibTeX
@article{Belalcazar2024,
  author = {Belalcazar, Andres and Heist, E Kevin},
  title = {Comparison of efficiency of PFA catheter designs by computer modeling},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2024},
  volume = {35},
  number = {12},
  pages = {2382--2393},
  doi = {10.1111/jce.16459}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche