Effects of parameters on radiofrequency guidewire ablation: In vitro and in vivo experiments

Zhongyin Zuo iD, Sainan Li iD, Fengqi Xuan, Jie Zhang, Zichen Liu, Shibei Zhang, Ming Liang, Zulu Wang iD

Journal of Cardiovascular Electrophysiology · 2023

DOI: 10.1111/jce.16066 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Zhongyin Zuo und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, General Hospital of Northern Theater Command of Jinzhou Medical University, Shenyang, China. Hauptthemen laut MEDLINE: Catheter Ablation, Radiofrequency Ablation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

A novel ablation technique with guidewire has emerged as a promising approach for mapping and ablation of arrhythmias originating from left ventricular summit. However, its biophysical characteristics have not been fully clarified. Methods and Results In the in vitro experiment, guidewire ablation (GA) was performed in vessel models of 1.17 and 2.24 mm to determine the maximum safety power. Then with the maximum safety power, the predictive value of generator impedance (GI) drop on lesion radius was explored. In the in vivo experiment, the feasibility of the maximum safety power and lesion formation was verified in the living swine. It was found that in both groups, the incidence of steam pops increased along with the raise of ablation power, and the maximum safety power was 10 W for the 1.17‐mm group and 15 W for the 2.24‐mm group. There was a strong linear correlation between GI drop and maximum lesion radius (in 1.17 mm‐10‐W group: r = .961; in 2.24 mm‐15‐W group: r = .918). In the in vivo experiment, besides ventricular fibrillation happened once, no other complications were observed, and lesions were found at both 48‐h and 8‐week groups.

Conclusions

The safety power of GA should be adjusted according to the diameter of the vessel. Besides, the GI drop can predict the lesion radius during GA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheter AblationRadiofrequency AblationSwineAnimalsHeart VentriclesArrhythmias, CardiacVentricular Fibrillation

Stichwörter der Autor:innen

guidewire ablationlesion radiusradiofrequency ablationsafety power

Autor:innen und Institutionen

  • Department of Cardiology, General Hospital of Northern Theater Command of Jinzhou Medical University, Shenyang, China.
  • Department of Cardiology, Shanghai Fourth People's Hospital Affiliated to Tongji University School of Medicine, Shanghai, China.
  • Department of Cardiology, General Hospital of Northern Theater Command of China Medical University, Shenyang, China.
  • Department of Cardiology, General Hospital of Northern Theater Command, Shenyang, China.
  • National Key Laboratory of Frigid Zone Cardiovascular Diseases, General Hospital of Northern Theater Command, Shenyang, China.

ORCID-Profile: Zhongyin Zuo, Sainan Li, Zulu Wang

Zitieren

APA

Zuo, Z., Li, S., Xuan, F., Zhang, J., Liu, Z., Zhang, S., Liang, M., & Wang, Z. (2023). Effects of parameters on radiofrequency guidewire ablation: In vitro and in vivo experiments. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16066

Vancouver

Zuo Z, Li S, Xuan F, Zhang J, Liu Z, Zhang S, et al. Effects of parameters on radiofrequency guidewire ablation: In vitro and in vivo experiments. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.16066

BibTeX
@article{Zuo2023,
  author = {Zuo, Zhongyin and Li, Sainan and Xuan, Fengqi and Zhang, Jie and Liu, Zichen and Zhang, Shibei and Liang, Ming and Wang, Zulu},
  title = {Effects of parameters on radiofrequency guidewire ablation: In vitro and in vivo experiments},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2023},
  doi = {10.1111/jce.16066}
}
Kostenlos starten →