Higher power achieves greater local impedance drop, shorter ablation time, and more transmural lesion formation in comparison to lower power in local impedance guided radiofrequency ablation of atrial fibrillation
Journal of Cardiovascular Electrophysiology · 2023
DOI: 10.1111/jce.16025 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Shu Yamashita und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, National Hospital Organization Disaster Medical Center, Tokyo, Japan. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
Since the local impedance (LI) of the ablation catheter reflects tissue characteristics, the efficacy of higher power (HP) compared to lower power (LP) in LI‐guided ablation may differ from other index‐guided ablations.
Objective
This study aimed to assess the efficacy of HP ablation in LI‐guided ablation of atrial fibrillation (AF).
Methods
A prospective observational study was conducted, enrolling patients undergoing de novo ablation for AF. Pulmonary vein isolation was performed using point‐by‐point ablation with a RHYTHMIA HDx TM Mapping System and an open‐irrigated ablation catheter with mini‐electrodes (IntellaNav MIFI OI). Ablation was stopped when the LI drop reached 30 ohms, three seconds after the LI plateaued, or when ablation time reached 30 s. To balance the baseline differences, a unique method was used in which the power was changed between HP (45 W to anterior wall/40 W to posterior wall) and LP (35 W/30 W) alternately for each adjacent point.
Results
A total of 551 ablations in 10 patients were analyzed (HP, n = 276; LP, n = 275). The maximum LI drop was significantly larger (HP: 28.3 ± 5.4 vs. LP: 24.8 ± 6.3 ohm), and the time to minimum LI was significantly shorter (HP: 15.0 ± 6.3 vs. LP: 19.3 ± 6.6 s) in the HP setting. The unipolar electrogram analysis of three patients revealed that the electrogram indicating transmural lesion formation was observed more frequently in the HP setting.
Conclusion
In LI‐guided ablation, the HP could achieve a larger LI drop and shorter time to minimum LI, which may result in more transmural lesion formation compared to a LP setting.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, National Hospital Organization Disaster Medical Center, Tokyo, Japan.
- Department of Cardiology, Kameda Medical Center, Chiba, Japan.
- Department of Cardiovascular Medicine, Tokyo Medical and Dental University, Tokyo, Japan.
ORCID-Profile: Shu Yamashita, Shinsuke Miyazaki, Tetsuo Sasano
Zitieren
APA
Yamashita, S., Mizukami, A., Ono, M., Hiroki, J., Miyakuni, S., Ueshima, D., Matsumura, A., Miyazaki, S., & Sasano, T. (2023). Higher power achieves greater local impedance drop, shorter ablation time, and more transmural lesion formation in comparison to lower power in local impedance guided radiofrequency ablation of atrial fibrillation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16025
Vancouver
Yamashita S, Mizukami A, Ono M, Hiroki J, Miyakuni S, Ueshima D, et al. Higher power achieves greater local impedance drop, shorter ablation time, and more transmural lesion formation in comparison to lower power in local impedance guided radiofrequency ablation of atrial fibrillation. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.16025