EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation

Péter Perge iD, Nikola Petrovic iD, Zoltán Salló iD, Katalin Piros iD, Vivien Klaudia Nagy iD, Pál Ábrahám iD, István Osztheimer iD, Béla Merkely iD, László Gellér iD, Nándor Szegedi iD

Europace · 2024

DOI: 10.1093/europace/euae260 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Europace, 2024. Verfasst von Péter Perge und 9 weiteren Autor:innen. Beteiligte Einrichtung: Heart and Vascular Center, Semmelweis University, Városmajor u. 68, 1122 Budapest, Hungary. Hauptthemen laut MEDLINE: Pulmonary Veins, Atrial Fibrillation, Electric Impedance, Catheter Ablation. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research, Development and Innovation Office of Hungary, European Union, National Academy of Scientist Education Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

The predictive role of local impedance (LI) drop in lesion formation using a novel contact force sensing ablation catheter was recently described. The purpose of our current study was to assess the temporal characteristics of LI drop during ablation and its correlation with acute lesion efficacy. Methods and results Point-by-point pulmonary vein isolation was performed. The efficacy of applications was determined by pacing along the circular ablation line and assessing loss of capture. Local impedance, contact force, and catheter position data with high resolution were analysed and compared in successful and unsuccessful applications. Five hundred and fifty-nine successful and 84 unsuccessful applications were analysed. The successful applications showed higher baseline LI (P < 0.001) and larger LI drop during ablation (P < 0.001, for all). In case of unsuccessful applications, after a moderate but significant drop from baseline to the 2 s time point (153 vs. 145 Ω, P < 0.001), LI did not change further (P = 0.99). Contradictorily, in case of successful applications, the LI significantly decreased further (baseline–2 s–10 s: 161–150–141 Ω, P < 0.001 for all). The optimal cut-point for the LI drop indicating unsuccessful application was <9 Ω at the 4-s time point [AUC = 0.73 (0.67–0.76), P < 0.001]. Failing to reach this cut-point predicted unsuccessful applications [OR 3.82 (2.34–6.25); P < 0.001].

Conclusion

A rapid and enduring drop of the LI may predict effective lesion formation, while slightly changing or unchanged LI is associated with unsuccessful applications. A moderate LI drop during the first 4 s of radiofrequency application predicts ineffective radiofrequency delivery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pulmonary VeinsAtrial FibrillationElectric ImpedanceCatheter AblationHumansMaleFemaleMiddle AgedAgedTreatment OutcomeTime FactorsCardiac Catheters

Stichwörter der Autor:innen

Atrial fibrillationCatheter ablationLocal impedancePulmonary vein isolation

Autor:innen und Institutionen

  • Heart and Vascular Center, Semmelweis University, Városmajor u. 68, 1122 Budapest, Hungary.
  • Boston Scientific International B.V., Kerkrade, The Netherlands.
  • Faculty of Mechanical Engineering, University of Belgrade, Belgrade, Serbia.

ORCID-Profile: Péter Perge, Nikola Petrovic, Zoltán Salló, Katalin Piros, Vivien Klaudia Nagy, Pál Ábrahám, István Osztheimer, Béla Merkely, László Gellér, Nándor Szegedi

Förderung

  • National Research, Development and Innovation Office of Hungary
  • European Union
  • National Academy of Scientist Education Program
  • National Biomedical Foundation
  • Hungarian Ministry of Culture and Innovation
  • MD-PhD Excellence Program of Semmelweis University
  • New National Excellence Program of the Ministry for Culture and Innovation
  • National Research, Development and Innovation Fund

Zitieren

APA

Perge, P., Petrovic, N., Salló, Z., Piros, K., Nagy, V. K., Ábrahám, P., Osztheimer, I., Merkely, B., Gellér, L., & Szegedi, N. (2024). Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation. Europace. https://doi.org/10.1093/europace/euae260

Vancouver

Perge P, Petrovic N, Salló Z, Piros K, Nagy VK, Ábrahám P, et al. Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation. Europace. 2024. doi:10.1093/europace/euae260

BibTeX
@article{Perge2024,
  author = {Perge, Péter and Petrovic, Nikola and Salló, Zoltán and Piros, Katalin and Nagy, Vivien Klaudia and Ábrahám, Pál and Osztheimer, István and Merkely, Béla and Gellér, László and Szegedi, Nándor},
  title = {Early rapid local impedance drop is associated with acute lesion efficacy during pulmonary vein isolation},
  journal = {Europace},
  year = {2024},
  doi = {10.1093/europace/euae260}
}