Distribution of excitation recoverable myocardium after radiofrequency ablation and its relation to energy application time and irrigation

Osamu Saitoh iD, Takumi Kasai, Kyogo Fuse, Masaomi Chinushi iD

Journal of Cardiovascular Electrophysiology · 2023

DOI: 10.1111/jce.15873 ↗

At a glance

Journal article in Journal of Cardiovascular Electrophysiology, 2023. Written by Osamu Saitoh and 3 co-authors. Institution: Cardiovascular Research of Graduate School of Health Sciences, Niigata University School of Medicine, Niigata, Japan. Major MEDLINE topics: Myocardium, Catheter Ablation. Cited 1 times so far (Crossref); 25 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Introduction

Radiofrequency (RF) catheter ablation induces excitation recoverable myocardium around durable core lesions, and its distribution may be different depending on energy delivery methods. Methods and Results In coronary perfusing porcine hearts, pacing threshold through the ventricle was measured using eight‐pole (1‐mm distance) needle electrodes vertically inserted into myocardium before, within 3 min after and 40 min after 40 W ablation with 10‐g catheter contact (Group 1: irrigation catheter for 15 s, Group 2: irrigation catheter for 40 s, Group 3: nonirrigation catheter for 15 s, Group 4: nonirrigation catheter for 40 s). Ablation was accomplished in all 12 ablations in Groups 1–3 whereas in 8/12 ablations in Group 4 because of high‐temperature rise. Within 3 min after ablation, 10.0 V pacing uncaptured electrodes were distributed from the surface to inside the myocardium, and its depth was deeper in 40 s than in 15 s ablation. 40 min after ablation, excitation recovery at one or more electrodes below the durable lesion was observed in all Groups. Excitation recovery electrodes were also observed on the surface in Group 1 but not the other Groups. Accordingly, the number of excitation‐recovered electrodes were larger in Group 1 than the other Groups.

Conclusions

Regardless of the ablation methods, excitation recoverable myocardium was present around 1.0 mm below the durable lesions. Lesions created by short application time using an irrigation catheter may have included large excitation recoverable myocardium soon after ablation because of the presence of reversible myocardium on well‐irrigated myocardial surfaces.

Subjects

MeSH terms (MEDLINE)

MyocardiumCatheter AblationAnimalsSwineHeartElectrodesHeart Ventricles

Author keywords

application timeirrigation catheterradiofrequency ablationrecoverable lesions

Authors and institutions

  • Cardiovascular Research of Graduate School of Health Sciences, Niigata University School of Medicine, Niigata, Japan.

ORCID profiles: Osamu Saitoh, Masaomi Chinushi

Cite

APA

Saitoh, O., Kasai, T., Fuse, K., & Chinushi, M. (2023). Distribution of excitation recoverable myocardium after radiofrequency ablation and its relation to energy application time and irrigation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.15873

Vancouver

Saitoh O, Kasai T, Fuse K, Chinushi M. Distribution of excitation recoverable myocardium after radiofrequency ablation and its relation to energy application time and irrigation. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.15873

BibTeX
@article{Saitoh2023,
  author = {Saitoh, Osamu and Kasai, Takumi and Fuse, Kyogo and Chinushi, Masaomi},
  title = {Distribution of excitation recoverable myocardium after radiofrequency ablation and its relation to energy application time and irrigation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2023},
  doi = {10.1111/jce.15873}
}
Start for free →