A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions
Medical Physics · 2025
DOI: 10.1002/mp.18112 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Ruizhe Hou und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, P.R. China. Hauptthemen laut MEDLINE: Phantoms, Imaging, Temperature, Catheter Ablation, Radiofrequency Ablation. Die Arbeit nennt 46 Literaturangaben. Gefördert von: National Natural Science Foundation of China, the Ministry of Science and Technology of China, the Interdisciplinary Program of Shanghai Jiao Tong University.
Abstract (Originalsprache)
Background
In catheter‐based radiofrequency ablation (RFA), energy is delivered to heterogeneous thin‐walled tissues to induce therapeutic heating. Variations in electrical and mechanical properties of tissue contents have a great effect on outcomes.
Purpose
The objective of this study is to develop models that replicate tissue heterogeneity and visualize ablation zones for effective evaluation and optimization.
Methods
A tunable thermochromic phantom formulation was developed, consisting of deionized water, acrylamide aqueous solution (Acr/Bis), thermochromic microcapsule powder, Tris‐HCl buffer, ammonium persulfate solution, and N , N , N ', N '‐tetramethylethylenediamine. The color‐change temperature threshold, electrical conductivity, and elastic modulus were characterized with different formulation compositions. A dual‐parameter inverse design method based on concentration‐property regression models was proposed to match the electrical and mechanical properties of tissues simultaneously, which was validated through RFA experiments on phantoms.
Results
The phantoms exhibited an immediate and irreversible color change from indigo to deep pink at 60°C, clearly delineating ablation margins. Electrical conductivity ranged from 0.1892 to 0.5958 S/m and elastic modulus from 6.70 to 138 kPa, effectively encompassing the characteristic properties of diverse lesion tissues. RFA experiments on phantoms mimicking porcine myocardium and chicken breast demonstrated strong agreement with ex vivo results. Ablation of heterogeneous phantoms mimicking atherosclerotic structures revealed significant differences compared to homogeneous models.
Conclusions
A tunable thermochromic phantom formulation was developed, exhibiting a precise color change at 60°C while replicating the electrical and mechanical properties of heterogeneous tissues, enabling direct visualization of ablation boundaries. The phantom offers a robust platform for evaluating catheter‐based RFA in clinically relevant complex tissues, supporting the development of personalized and optimized treatment strategies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, P.R. China.
Förderung
- National Natural Science Foundation of China
- the Ministry of Science and Technology of China
- the Interdisciplinary Program of Shanghai Jiao Tong University
Zitieren
APA
Hou, R., Han, Q., Zou, J., Zhao, S., & Zhang, A. (2025). A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions. Medical Physics. https://doi.org/10.1002/mp.18112
Vancouver
Hou R, Han Q, Zou J, Zhao S, Zhang A. A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions. Medical Physics. 2025. doi:10.1002/mp.18112
BibTeX
@article{Hou2025,
author = {Hou, Ruizhe and Han, Quanshu and Zou, Jincheng and Zhao, Shiqing and Zhang, Aili},
title = {A thermochromic tissue‐mimicking phantom for catheter‐based radiofrequency ablation of heterogenous lesions},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.18112}
}
EUVIMED