Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i
International Journal of Laboratory Hematology · 2026-08 · 48 (4) · 784-794
DOI: 10.1111/ijlh.70107 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift International Journal of Laboratory Hematology, online erschienen am 10.04.2026. Verfasst von Hyeongyu Lee und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea. Hauptthemen laut MEDLINE: Hematologic Tests. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
The Atellica HEMA 580 is a novel haematology analyser providing eight‐part leukocyte differential and optical platelet (PLT) counting. We compared its analytical and clinical performance with that of the ADVIA 2120i.
Methods
A total of 796 residual whole blood specimens were analysed. Carryover, linearity and precision were assessed. Inter‐instrument comparison was performed for complete blood counts and five‐part differential (5‐Diff) parameters using Passing–Bablok regression and Bland–Altman analyses. Additional regression against manual microscopy was conducted for flagged samples, including PLT, reticulocytes and common 5‐Diff parameters.
Results
Carryover was < 1% for all parameters. Linearity was excellent ( R 2 ≥ 0.99). Precision was within acceptable limits; coefficient of variant was < 5% for most parameters, except optical PLT, lymphocytes, monocytes, eosinophils and nucleated red blood cells under specific conditions. Inter‐instrument comparisons revealed remarkably high correlations ( r ≥ 0.9) for most parameters, whereas mean corpuscular haemoglobin concentration (MCHC) ( r = 0.429) and basophils ( r = 0.627) showed weaker concordance. Passing–Bablok regression indicated minimal systematic or proportional bias for most parameters. Bland–Altman analysis showed that > 95% of results were within the 95% limits of agreement, except white blood cells, mean corpuscular volume and PLT (94.7%–94.8%). Comparison with microscopy confirmed good concordance for PLT and reticulocytes; however, lymphocytes and monocytes had proportional and systematic biases.
Conclusion
The Atellica HEMA 580 demonstrated reliable analytical performance and strong agreement with the ADVIA 2120i. This supports its suitability as a routine haematology analyser, providing accurate and clinically comparable results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea.
ORCID-Profile: Jooyoung Cho
Zitieren
APA
Lee, H., Cho, J., Yoo, G., & Kim, J. (2026). Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i. International Journal of Laboratory Hematology, 48(4), 784-794. https://doi.org/10.1111/ijlh.70107
Vancouver
Lee H, Cho J, Yoo G, Kim J. Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i. International Journal of Laboratory Hematology. 2026;48(4):784-794. doi:10.1111/ijlh.70107
BibTeX
@article{Lee2026,
author = {Lee, Hyeongyu and Cho, Jooyoung and Yoo, Gilsung and Kim, Juwon},
title = {Analytical and Clinical Performance of the Atellica HEMA 580 Haematology Analyzer: A Comparison With the ADVIA 2120i},
journal = {International Journal of Laboratory Hematology},
year = {2026},
volume = {48},
number = {4},
pages = {784--794},
doi = {10.1111/ijlh.70107}
}
EUVIMED