Integrated ESR and PLT ‐ H Measurement Using the BC ‐6800 Plus Hematology Analyzer: A Comprehensive Analytical Evaluation

Meilin Chen, Jinghai Yan, Manman Ye, Shihong Zhang, Xiaohe Zheng iD

Journal of Clinical Laboratory Analysis · 2026

DOI: 10.1002/jcla.70185 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, 2026. Verfasst von Meilin Chen und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-Sen University, Guangzhou, China. Hauptthemen laut MEDLINE: Blood Sedimentation. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wu Jieping Medical Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Complete blood count (CBC) and erythrocyte sedimentation rate (ESR) are essential tests in clinical practice. The Mindray BC‐6800 Plus integrates CBC and ESR analysis through its Easy‐W ESR module and a hybrid platelet channel (PLT‐H) designed to improve platelet quantification, especially under interference.

Methods

A prospective analytical study was conducted following CLSI guidelines (EP05‐A3, EP06‐A, EP09c). A total of 460 venous samples were analyzed across multiple modes (Auto, Open‐Vial, Predilute). PLT‐H was compared with PLT‐O and reference values; ESR with the BC‐760CS, which shares the same ESR module validated against Westergren. Interference testing included samples with low MCV (< 75 fL) and high MPV (> 10 fL).

Results

PLT‐H background was consistently < 5 × 10 9 /L. Precision met predefined limits (CV ≤ 4% whole blood, ≤ 8% predilute; ESR SD ≤ 2 mm/h at ≤ 20 mm/h, CV ≤ 10% at > 20 mm/h). Carryover was ≤ 0.2%. PLT‐H showed excellent linearity to 5000 × 10 9 /L ( r = 0.999) and tighter bias distribution than PLT‐O. Agreement was strong for ESR between BC‐6800 Plus and BC‐760CS ( r = 0.989) and for PLT‐H versus PLT‐O ( r = 0.992). PLT‐H maintained robustness in interference‐prone samples ( r = 0.992).

Conclusion

The BC‐6800 Plus demonstrated excellent analytical performance for ESR and PLT‐H, delivering accurate, reproducible, and interference‐resistant results. It supports efficient, high‐quality testing in modern laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood SedimentationHumansPlatelet CountReproducibility of ResultsBlood Cell CountProspective Studies

Stichwörter der Autor:innen

Mindray BC‐6800 plusanalytical performanceerythrocyte sedimentation rate (ESR)hematology analyzerplatelet hybrid count (PLT‐H)

Autor:innen und Institutionen

  • Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-Sen University, Guangzhou, China.

ORCID-Profile: Xiaohe Zheng

Förderung

  • Wu Jieping Medical Foundation

Zitieren

APA

Chen, M., Yan, J., Ye, M., Zhang, S., & Zheng, X. (2026). Integrated ESR and PLT ‐ H Measurement Using the BC ‐6800 Plus Hematology Analyzer: A Comprehensive Analytical Evaluation. Journal of Clinical Laboratory Analysis. https://doi.org/10.1002/jcla.70185

Vancouver

Chen M, Yan J, Ye M, Zhang S, Zheng X. Integrated ESR and PLT ‐ H Measurement Using the BC ‐6800 Plus Hematology Analyzer: A Comprehensive Analytical Evaluation. Journal of Clinical Laboratory Analysis. 2026. doi:10.1002/jcla.70185

BibTeX
@article{Chen2026,
  author = {Chen, Meilin and Yan, Jinghai and Ye, Manman and Zhang, Shihong and Zheng, Xiaohe},
  title = {Integrated ESR and PLT ‐ H Measurement Using the BC ‐6800 Plus Hematology Analyzer: A Comprehensive Analytical Evaluation},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2026},
  doi = {10.1002/jcla.70185}
}
Kostenlos starten →