Electrophysiological Recording With Activated Iridium Oxide: Effective Electrode Area, Not Total Impedance, Determines Performance
European Journal of Neuroscience · 2026-07 · 64 (2)
DOI: 10.1111/ejn.70640 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Neuroscience, online erschienen am 20.07.2026. Verfasst von Alexander R Harris und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Melbourne, Melbourne, Victoria, Australia. Hauptthemen laut MEDLINE: Iridium, Neurons. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Australian Research Council Centre of Excellence Scheme.
Abstract (Originalsprache)
Platinum electrodes are used in many bionic devices; however, they have limited charge injection capacity (CIC) and signal‐to‐noise ratio (SNR), restricting performance. Iridium oxide has substantially larger CIC and reduced impedance compared to platinum. While these types of electrochemical measurements are typically used to infer improved electrophysiological performance, there are no studies demonstrating this correlation. This article compares the electrochemical and acute electrophysiological recording performance of iridium and electrochemically activated IrOx. Increased iridium activation reduced total impedance at low‐intermediate frequencies and steady‐state electroactive area but had no impact on linear diffusion electroactive area. The total impedance had a poor correlation with background noise, SNR and spike count. The results contribute to the previous literature indicating linear diffusion electroactive area determines the number of neurons within recording distance of the electrode, impacting SNR and spike count. Measurements of impedance are mainly of benefit as an indirect measure of linear diffusion electroactive area, in which case the measurement and analysis must be performed appropriately. A wide range of IrOx activation states can be produced with increased CIC, without concern of neural recording performance impacts. However, higher oxidation states resulting in electrode dissolution or increase in electrode resistance may limit IrOx chronic performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Melbourne, Melbourne, Victoria, Australia.
- Higher Education College, Chisholm Institute, Dandenong, Victoria, Australia.
- ISN Psychology, Institute for Social Neuroscience, Ivanhoe, Victoria, Australia.
- School of Psychology and Public Health, La Trobe University, Bundoora, Victoria, Australia.
ORCID-Profile: Alexander R Harris
Förderung
- Australian Research Council Centre of Excellence Scheme
Zitieren
APA
Harris, A. R., Allitt, B. J., & Paolini, A. G. (2026). Electrophysiological Recording With Activated Iridium Oxide: Effective Electrode Area, Not Total Impedance, Determines Performance. European Journal of Neuroscience, 64(2). https://doi.org/10.1111/ejn.70640
Vancouver
Harris AR, Allitt BJ, Paolini AG. Electrophysiological Recording With Activated Iridium Oxide: Effective Electrode Area, Not Total Impedance, Determines Performance. European Journal of Neuroscience. 2026;64(2). doi:10.1111/ejn.70640
BibTeX
@article{Harris2026,
author = {Harris, Alexander R and Allitt, Ben J and Paolini, Antonio G},
title = {Electrophysiological Recording With Activated Iridium Oxide: Effective Electrode Area, Not Total Impedance, Determines Performance},
journal = {European Journal of Neuroscience},
year = {2026},
volume = {64},
number = {2},
doi = {10.1111/ejn.70640}
}
EUVIMED