Core Laboratory Versus Center‐Reported Echocardiographic Assessment of the Native and Bioprosthetic Aortic Valve

Bart J J Velders iD, Michiel D Vriesendorp, Neil J Weissman, Joseph F Sabik, Michael J Reardon, Francois Dagenais, Michael G Moront, Vivek Rao, Shinichi Fukuhara, Ralf Günzinger, Wouter J van Leeuwen, W Morris Brown and 3 more

Echocardiography · 2024

DOI: 10.1111/echo.70047 ↗

At a glance

Multizentrische Studie in Echocardiography, 2024. Written by Bart J J Velders and 14 co-authors. Institution: Cardiothoracic Surgery, Leiden University Medical Centre, Leiden, the Netherlands. Major MEDLINE topics: Bioprosthesis, Heart Valve Prosthesis, Aortic Valve, Echocardiography. Cited 3 times so far (Crossref); 23 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Medtronic.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Insights into quantitative differences between core laboratory and center‐reported echocardiographic assessment of the native and bioprosthetic aortic valve are lacking. We aimed to explore clinically relevant differences between these evaluations.

Methods

Data were used from the PERIcardial SurGical AOrtic Valve ReplacemeNt (PERIGON) Pivotal Trial for the Avalus valve. In this trial, patients with an indication for surgical aortic valve replacement (SAVR) due to aortic stenosis or regurgitation (AR) were enrolled. Serial echocardiographic examinations were performed at each center and blindly reanalyzed by an independent echocardiographic core laboratory (ECL). For the bioprosthetic valve analysis, postoperative data throughout the 5‐year follow‐up were pooled. Differences between the ECL and the centers in continuous parameters were quantified in mean differences and intraclass correlation coefficients (ICCs). Agreement on AR, paravalvular leak (PVL), and prosthesis‐patient mismatch (PPM) classification was investigated using Cohen's kappa coefficients.

Results

The analysis on the native aortic valve was performed on 1118 echocardiograms. The relative mean difference was largest for the left ventricular outflow tract (LVOT) area, followed by stroke volume and effective orifice area (index), with center‐reported values being 11%–7% higher. High ICCs of around 0.90 were observed for the parameters peak aortic jet velocity, mean pressure gradient, and the velocity‐time integral across the aortic valve. Over 5000 echocardiograms were available for the bioprosthetic valve analysis. Therein, comparable results were observed. The kappa coefficient was 0.59 (95% confidence interval [CI] 0.56, 0.63) for agreement on native AR, 0.28 (95% CI 0.18, 0.37) for PVL, and 0.42 (95% CI 0.40, 0.44) for PPM.

Conclusions

There is high agreement between the ECL and clinical centers on continuous‐wave Doppler‐related measurements. In contrast, agreement is low for parameters that involve measurement of the LVOT diameter. These results provide important context for the interpretation of aortic valve performance in studies that lack central ECL evaluation. Trial Registration ClinicalTrials.gov identifier: NCT02088554

Subjects

MeSH terms (MEDLINE)

BioprosthesisHeart Valve ProsthesisAortic ValveEchocardiographyHumansMaleFemaleAgedReproducibility of ResultsAortic Valve StenosisTreatment OutcomeAortic Valve Insufficiency

Author keywords

bioprosthetic aortic valvecore laboratoryechocardiographyinterobserver variabilitynative aortic valve

Authors and institutions

  • Cardiothoracic Surgery, Leiden University Medical Centre, Leiden, the Netherlands.
  • Cardiovascular Core Laboratories, MedStar Health Research Institute, and Georgetown University, Washington, District of Columbia, USA.
  • Surgery, University Hospitals Cleveland Medical Center and Case Western Reserve University School of Medicine, Cleveland, Ohio, USA.
  • Department of Cardiothoracic Surgery, Houston Methodist DeBakey Heart and Vascular Center, Houston, Texas, USA.
  • Division of Cardiac Surgery, Quebec Heart and Lung Institute, Quebec, Quebec, Canada.
  • Cardiothoracic Surgery, ProMedica Toledo Hospital, Toledo, Ohio, USA.

ORCID profiles: Bart J J Velders

Funding

  • Medtronic

Cite

APA

Velders, B. J. J., Vriesendorp, M. D., Weissman, N. J., Sabik, J. F., Reardon, M. J., Dagenais, F., Moront, M. G., Rao, V., Fukuhara, S., Günzinger, R., van Leeuwen, W. J., Brown, W. M., Groenwold, R. H. H., Klautz, R. J. M., & Asch, F. M. (2024). Core Laboratory Versus Center‐Reported Echocardiographic Assessment of the Native and Bioprosthetic Aortic Valve. Echocardiography. https://doi.org/10.1111/echo.70047

Vancouver

Velders BJJ, Vriesendorp MD, Weissman NJ, Sabik JF, Reardon MJ, Dagenais F, et al. Core Laboratory Versus Center‐Reported Echocardiographic Assessment of the Native and Bioprosthetic Aortic Valve. Echocardiography. 2024. doi:10.1111/echo.70047

BibTeX
@article{Velders2024,
  author = {Velders, Bart J J and Vriesendorp, Michiel D and Weissman, Neil J and Sabik, Joseph F and Reardon, Michael J and Dagenais, Francois and Moront, Michael G and Rao, Vivek and Fukuhara, Shinichi and Günzinger, Ralf and van Leeuwen, Wouter J and Brown, W Morris and Groenwold, Rolf H H and Klautz, Robert J M and Asch, Federico M},
  title = {Core Laboratory Versus Center‐Reported Echocardiographic Assessment of the Native and Bioprosthetic Aortic Valve},
  journal = {Echocardiography},
  year = {2024},
  doi = {10.1111/echo.70047}
}
Start for free →