The Effect of Crown Height and Retentive Height of Abutment on the Fracture Resistance of Fatigued Monolithic Zirconia Anterior Implant Crowns—An In Vitro Study
Clinical Oral Implants Research · 2026
DOI: 10.1111/clr.70130 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2026. Verfasst von Esha Zahid und 5 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, Dentistry and Health Sciences, Melbourne Dental School, The University of Melbourne, Carlton, Australia. Hauptthemen laut MEDLINE: Crowns, Zirconium, Dental Restoration Failure, Dental Abutments. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Straumann Australia and New Zealand, Melbourne Dental School Internal Research Grant, University of Melbourne, Australia, Mac.Rob Foundation.
Abstract (Originalsprache)
Objectives
This laboratory study investigated the effect of two different retentive heights of Ti‐base abutments (3.5 and 5.5 mm) and crown heights (13 and 16 mm) on the load at failure and survival of monolithic zirconia implant‐supported crowns replacing a maxillary central incisor. Materials and Method Forty implant‐supported crowns were divided into four groups ( n = 10 each): short crown/long abutment (SCLA), short crown/short abutment (SCSA), long crown/long abutment (LCLA), and long crown/short abutment (LCSA). Specimens underwent thermal cycling and incremental mechanical cyclic loading starting at 450 N increasing by 50 N until failure. Load and cycles at failure, Weibull analysis, mode of failure, and fractographic assessments were performed.
Results
Mean load at failure was highest for the SCSA group (536.9 N ± 78.4 N), then SCLA (493.4 ± 53.2 N) and LCLA (491.3 ± 40.7 N), and lowest for the LCSA group (453.7 N ± 2.8 N). The differences were statistically significant ( p = 0.035). Cycles at failure showed significant differences among groups ( p < 0.001), with SCSA demonstrating the greatest cycles at failure (72,492 ± 44,551 cycles), then SCLA (43,630 ± 27,579 cycles) and LCLA (37,706 ± 18,826 cycles), and LCSA the lowest (8399 ± 7308 cycles). Short crowns showed abutment/screw fractures, while longer crowns had more crown fractures.
Conclusion
Crown height significantly influenced load at failure and survival, with shorter crowns performing better. The retentive height of the Ti‐base abutment had less impact; however, shorter abutments provided better mechanical performance when paired with shorter crowns, suggesting careful consideration for anterior restorations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Faculty of Medicine, Dentistry and Health Sciences, Melbourne Dental School, The University of Melbourne, Carlton, Australia.
ORCID-Profile: Esha Zahid, Jaafar Abduo
Förderung
- Straumann Australia and New Zealand
- Melbourne Dental School Internal Research Grant, University of Melbourne, Australia
- Mac.Rob Foundation
Zitieren
APA
Zahid, E., Abduo, J., Gergely, A., Judge, R., Perez-Rodriguez, C., & Palamara, J. (2026). The Effect of Crown Height and Retentive Height of Abutment on the Fracture Resistance of Fatigued Monolithic Zirconia Anterior Implant Crowns—An In Vitro Study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.70130
Vancouver
Zahid E, Abduo J, Gergely A, Judge R, Perez-Rodriguez C, Palamara J. The Effect of Crown Height and Retentive Height of Abutment on the Fracture Resistance of Fatigued Monolithic Zirconia Anterior Implant Crowns—An In Vitro Study. Clinical Oral Implants Research. 2026. doi:10.1111/clr.70130
BibTeX
@article{Zahid2026,
author = {Zahid, Esha and Abduo, Jaafar and Gergely, Attila and Judge, Roy and Perez-Rodriguez, Carolina and Palamara, Joseph},
title = {The Effect of Crown Height and Retentive Height of Abutment on the Fracture Resistance of Fatigued Monolithic Zirconia Anterior Implant Crowns—An In Vitro Study},
journal = {Clinical Oral Implants Research},
year = {2026},
doi = {10.1111/clr.70130}
}
EUVIMED