Accuracy (Trueness and Precision) of the Calibrated Intraoral Scan Protocol: An In Vitro Comparison of Three Scan Calibrator Designs In an Edentulous Implant Model

Shengtao Yang iD, Chenyi Zhang, Yan Bao, Shuo Wang, Lei Chi, Li Yue iD, Quan Yuan iD, Junying Li

Clinical Oral Implants Research · 2026-06 · 37 (6) · 747-759

DOI: 10.1111/clr.70117 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, online erschienen am 14.03.2026. Verfasst von Shengtao Yang und 7 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Dental Tec…. Hauptthemen laut MEDLINE: Dental Implants, Dental Impression Technique, Imaging, Three-Dimensional. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: West China Hospital of Stomatology, Sichuan University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To introduce a universal scan calibrator (SC) and to evaluate its accuracy in the calibrated intraoral scan protocol (CISP) by comparison with two previously published SCs. Materials and Methods Three SCs were evaluated on an edentulous implant model containing six scan bodies (SBs) positioned at the screw‐retained abutment (SRA) level: (1) a universal SC with a C‐shaped window (SC‐CW), (2) an SC covering only the alveolar ridge (SC‐R), and (3) an SC limited to the palatal region (SC‐P). Each SC was attached to the model and scanned 10 times using an intraoral scanner (IOS). The intraoral scans were subsequently corrected using the corresponding desktop scans of the SCs as reference data. A desktop scan of the model served as the reference. The test scans were superimposed onto the reference scan by: (1) using all SBs to evaluate the overall fit, and (2) using the first and second SBs to simulate the virtual Sheffield test. Linear and angular deviations of SRAs were measured using Python scripts.

Results

When aligned by all SBs, all three groups exhibited significant improvements in trueness and linear precision following correction. In the virtual Sheffield test, no statistically significant differences were observed in post‐correction trueness or angular precision at the ending SRA6. All three groups achieved clinically acceptable trueness and precision values following correction.

Conclusions

All three SCs yielded clinically acceptable impressions after correction under in vitro conditions. The novel SC represents a promising approach for CISP; however, further in vivo clinical evidence is required.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental ImplantsDental Impression TechniqueImaging, Three-DimensionalHumansCalibrationModels, DentalIn Vitro TechniquesComputer-Aided DesignDental Abutments

Stichwörter der Autor:innen

accuracycalibrated intraoral scan protocolcomplete‐arch implant impressiondigital dentistryintraoral scanscan calibrator

Substanzen

Dental Implants

Autor:innen und Institutionen

  • State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Dental Technology, West China Hospital of Stomatology, Sichuan University, Cheng…
  • State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Oral Implantology, West China Hospital of Stomatology, Sichuan University, Cheng…
  • Department of Biologic and Materials Sciences & Prosthodontics, University of Michigan School of Dentistry, Ann Arbor, Michigan, USA.

ORCID-Profile: Shengtao Yang, Li Yue, Quan Yuan

Förderung

  • West China Hospital of Stomatology
  • Sichuan University

Zitieren

APA

Yang, S., Zhang, C., Bao, Y., Wang, S., Chi, L., Yue, L., Yuan, Q., & Li, J. (2026). Accuracy (Trueness and Precision) of the Calibrated Intraoral Scan Protocol: An In Vitro Comparison of Three Scan Calibrator Designs In an Edentulous Implant Model. Clinical Oral Implants Research, 37(6), 747-759. https://doi.org/10.1111/clr.70117

Vancouver

Yang S, Zhang C, Bao Y, Wang S, Chi L, Yue L, et al. Accuracy (Trueness and Precision) of the Calibrated Intraoral Scan Protocol: An In Vitro Comparison of Three Scan Calibrator Designs In an Edentulous Implant Model. Clinical Oral Implants Research. 2026;37(6):747-759. doi:10.1111/clr.70117

BibTeX
@article{Yang2026,
  author = {Yang, Shengtao and Zhang, Chenyi and Bao, Yan and Wang, Shuo and Chi, Lei and Yue, Li and Yuan, Quan and Li, Junying},
  title = {Accuracy (Trueness and Precision) of the Calibrated Intraoral Scan Protocol: An In Vitro Comparison of Three Scan Calibrator Designs In an Edentulous Implant Model},
  journal = {Clinical Oral Implants Research},
  year = {2026},
  volume = {37},
  number = {6},
  pages = {747--759},
  doi = {10.1111/clr.70117}
}
Kostenlos starten →