Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study
Journal of Dentistry · 2026
DOI: 10.1016/j.jdent.2026.106983 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Dentistry, 2026. Verfasst von Siyu Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Dental Tec…. Hauptthemen laut MEDLINE: Dental Implants, Dental Arch, Computer-Aided Design, Imaging, Three-Dimensional. Die Arbeit nennt 48 Literaturangaben. Gefördert von: West China Hospital, Sichuan University, Sichuan University, Sichuan University West China Hospital of Stomatology.
Abstract (Originalsprache)
OBJECTIVE: To evaluate the effect of four scan calibrator (SC) morphologies on the accuracy of calibrated complete-arch implant intraoral scans. METHODS: A study model featuring six scan bodies (SBs) at the screw-retained abutment (SRA) level was fabricated, and reference data were obtained by scanning the model with an industrial scanner. A universal SC was manufactured in four morphologies: (1) tubular (SC-TU), (2) crumpled (SC-CR), (3) spherical (SC-SP), and (4) polyhedral (SC-PO). Each SC was attached to the model, and the resulting model-SC assembly was scanned ten times (n = 10) using an intraoral scanner (IOS) to obtain uncalibrated scans. These uncalibrated datasets were then calibrated using the corresponding desktop scans of the SCs. Both the uncalibrated and calibrated scans were registered against the reference scan via two alignment strategies: (1) using all SBs to assess the overall fit, and (2) using only the first and second SBs to simulate the Sheffield test. Linear deviations between the central points of SRAs were calculated using a custom Python script. RESULTS: When aligned using all SBs, trueness was comparable among the four SC-assisted groups, whereas the SC-CR group exhibited significantly lower post-calibration precision (40.11 ± 35.21 μm) than the SC-SP (29.63 ± 17.37 μm) and SC-PO (28.08 ± 13.48 μm) groups. In the virtual Sheffield test, the SC-SP (86.30 ± 34.25 µm) group demonstrated significantly higher trueness than the SC-CR group (162.06 ± 62.45 µm) following calibration at the distal SRA6. Furthermore, the SC-PO (66.21 ± 29.96 μm) group exhibited significantly higher post-calibration precision than the other three SC groups at SRA6. CONCLUSIONS: SC morphologies influence the accuracy of calibrated complete-arch implant intraoral scans. SC-PO demonstrated the highest post-calibration precision. CLINICAL SIGNIFICANCE: The proposed SC morphologies, except for SC-CR, may help correct IOS stitching errors. The polyhedral morphology may represent a favorable design choice for SC fabrication.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Dental Technology, West China Hospital of Stomatology, Sichuan University, Cheng…
- Department of Biologic and Materials Sciences & Prosthodontics, University of Michigan School of Dentistry, Ann Arbor, MI, USA.
- State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Oral Implantology, West China Hospital of Stomatology, Sichuan University, Cheng…
Förderung
- West China Hospital, Sichuan University
- Sichuan University
- Sichuan University West China Hospital of Stomatology
Zitieren
APA
Wang, S., Yang, S., Gu, Y., Yue, L., Li, J., & Yuan, Q. (2026). Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study. Journal of Dentistry. https://doi.org/10.1016/j.jdent.2026.106983
Vancouver
Wang S, Yang S, Gu Y, Yue L, Li J, Yuan Q. Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study. Journal of Dentistry. 2026. doi:10.1016/j.jdent.2026.106983
BibTeX
@article{Wang2026,
author = {Wang, Siyu and Yang, Shengtao and Gu, Yiping and Yue, Li and Li, Junying and Yuan, Quan},
title = {Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study},
journal = {Journal of Dentistry},
year = {2026},
doi = {10.1016/j.jdent.2026.106983}
}
EUVIMED