Randomisierte kontrollierte Studie in der Zeitschrift Clinical Oral Implants Research, 2026. Verfasst von Marc Balmer und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Reconstructive Dentistry, Center for Dental Medicine Zurich, University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Zirconium, Titanium, Dental Implants, Dental Implants, Single-Tooth. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: International Team for Implantology.
Abstract (Originalsprache)
Objectives
To evaluate the clinical performance of two‐piece zirconia implants with screw‐retained abutments compared to titanium implants after 1 year of loading. Materials and Methods In this multicenter, prospective, randomized clinical trial, 61 two‐piece zirconia and 61 titanium implants were placed and restored with single crowns. Evaluations were performed at implant placement, crown delivery, and 1‐year post‐loading. Peri‐implant Marginal Bone Loss (MBL), survival rate, early wound healing index, and soft tissue parameters were assessed. Intergroup comparisons of continuous outcomes were performed using Linear Mixed‐Effects Models accounting for center and potential confounders. Categorical variables were analyzed using the chi‐square or Fisher's exact test. Survival was analyzed using Kaplan–Meier estimates. Significance was set at p < 0.05.
Results
No significant differences were found between the two implant types. Mean MBL from implantation to crown insertion was 1.10 ± 0.78 mm for titanium and 0.94 ± 0.67 mm for zirconia implants. No significant additional bone loss occurred over the subsequent year, with changes of 0.07 ± 0.55 mm and 0.08 ± 0.51 mm for titanium and zirconia, respectively. After 1 year, zirconia implants showed a 100% survival rate, while titanium implants showed 96.5% with two failures. At 1 year, differences in probing depths, plaque accumulation, and Papilla Bleeding Index were not statistically significant.
Conclusion
After 1 year of loading, no statistically significant differences in MBL, implant survival, or peri‐implant health were found between zirconia and titanium implants, indicating no clinical superiority. Zirconia implants may therefore be considered a viable alternative in single‐tooth implant restorations. Trial Registration The study is registered at the German Clinical Trial Register (No. DRKS 00013209) as well as at the Federal Office of Public Health's (FOPH) portal for human research in Switzerland (kofam.ch)
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Clinic of Reconstructive Dentistry, Center for Dental Medicine Zurich, University of Zurich, Zurich, Switzerland.
- Department of Oral Surgery and Orthodontics, University Clinic for Dental Medicine & Oral Health, Medical University of Graz, Graz, Austria.
- Department of Restorative Dentistry, Periodontology and Prosthodontics, University Clinic for Dental Medicine & Oral Health, Medical University of Graz, Graz, Austria.
- Department of Oral Surgery, University Center for Dental Medicine in Basel, University of Basel, Basel, Switzerland.
ORCID-Profile: Marc Balmer, Michael Payer, Anke Steinwender, Valentin Herber, Ronald E Jung, Sebastian Kühl
Förderung
- International Team for Implantology
Zitieren
APA
Balmer, M., Payer, M., Steinwender, A., Herber, V., Jung, R. E., & Kühl, S. (2026). One‐Year Analysis of Clinical and Radiological Outcomes of Two‐Piece Zirconia Compared to Titanium Implants: A Multicenter Prospective Randomized Clinical Trial. Clinical Oral Implants Research. https://doi.org/10.1111/clr.70094
Vancouver
Balmer M, Payer M, Steinwender A, Herber V, Jung RE, Kühl S. One‐Year Analysis of Clinical and Radiological Outcomes of Two‐Piece Zirconia Compared to Titanium Implants: A Multicenter Prospective Randomized Clinical Trial. Clinical Oral Implants Research. 2026. doi:10.1111/clr.70094
BibTeX
@article{Balmer2026,
author = {Balmer, Marc and Payer, Michael and Steinwender, Anke and Herber, Valentin and Jung, Ronald E and Kühl, Sebastian},
title = {One‐Year Analysis of Clinical and Radiological Outcomes of Two‐Piece Zirconia Compared to Titanium Implants: A Multicenter Prospective Randomized Clinical Trial},
journal = {Clinical Oral Implants Research},
year = {2026},
doi = {10.1111/clr.70094}
}
EUVIMED