Fachartikel in der Zeitschrift Clinical Oral Implants Research, online erschienen am 20.12.2024. Verfasst von Matheus L Oliveira und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Brazil. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Artifacts, Dental Implants, Dental Materials. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Fundação de Amparo à Pesquisa do Estado de São Paulo, Swiss Association of Dentomaxillofacial Radiology (SGDMFR/SADMFR).
Abstract (Originalsprache)
Objectives
To evaluate the influence of different dental implant materials within the exomass on the image quality in cone‐beam computed tomography (CBCT). Material and Methods Five pig jaws were scanned using four CBCT devices, first without any dental implants, followed by scans with three dental implants of the same material—pure titanium, titanium‐zirconium alloy, and zirconium dioxide. Two fields of view (FOVs) were used for each device to position the implants in the exomass of a small FOV and within a large FOV. Voxel values were obtained from tubes containing a radiopaque solution to calculate mean voxel value (MVV), voxel value inhomogeneity (VVI), and image noise (IN), which were compared across implant materials and FOVs using repeated measures analysis of variance ( α = 0.05). Three observers independently scored image quality using a 5‐point scale.
Results
In general, MVV remained unaffected, except for a significant increase in the X800 device when titanium or titanium‐zirconium alloy implants were in the exomass of a small FOV ( p ≤ 0.05). A trend of increased VVI was observed when implants were in the exomass of a small FOV, with a greater effect for zirconium dioxide, followed by titanium‐zirconium alloy and titanium. IN was higher when implants were in the exomass, especially for zirconium dioxide implants ( p ≤ 0.05). Image quality perception was consistent overall, though zirconium dioxide implants in both FOVs resulted in diminished quality.
Conclusions
The presence of implants in the exomass can negatively affect CBCT image quality, with zirconium dioxide having the greatest impact.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Brazil.
- Department of Oral Health & Medicine, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
- Dental Imaging, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
- Department of Oral Surgery, University, Center for Dental Medicine, University of Basel, Basel, Switzerland.
ORCID-Profile: Matheus L Oliveira, Dorothea Dagassan-Berndt, Matheus Sampaio-Oliveira, Michelle Simonek, Sebastian Kühl, Michael M Bornstein
Förderung
- Fundação de Amparo à Pesquisa do Estado de São Paulo
- Swiss Association of Dentomaxillofacial Radiology (SGDMFR/SADMFR)
Zitieren
APA
Oliveira, M. L., Dagassan-Berndt, D., Sampaio-Oliveira, M., Simonek, M., Kühl, S., & Bornstein, M. M. (2025). Evaluation of Exomass‐Related Artefacts Caused by Dental Implants of Different Materials in Cone‐Beam Computed Tomography Scans: An Ex Vivo Study. Clinical Oral Implants Research, 36(4), 449-459. https://doi.org/10.1111/clr.14394
Vancouver
Oliveira ML, Dagassan-Berndt D, Sampaio-Oliveira M, Simonek M, Kühl S, Bornstein MM. Evaluation of Exomass‐Related Artefacts Caused by Dental Implants of Different Materials in Cone‐Beam Computed Tomography Scans: An Ex Vivo Study. Clinical Oral Implants Research. 2025;36(4):449-459. doi:10.1111/clr.14394
BibTeX
@article{Oliveira2025,
author = {Oliveira, Matheus L and Dagassan-Berndt, Dorothea and Sampaio-Oliveira, Matheus and Simonek, Michelle and Kühl, Sebastian and Bornstein, Michael M},
title = {Evaluation of Exomass‐Related Artefacts Caused by Dental Implants of Different Materials in Cone‐Beam Computed Tomography Scans: An Ex Vivo Study},
journal = {Clinical Oral Implants Research},
year = {2025},
volume = {36},
number = {4},
pages = {449--459},
doi = {10.1111/clr.14394}
}
EUVIMED