Metal artifact reduction improves cone-beam CT diagnosis of peri-implant bone defects regardless of adjacent implant position in endomass or exomass
Dentomaxillofacial Radiology · 2026
DOI: 10.1093/dmfr/twag025 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Lucas Santana-Santos und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Diagnosis, Piracicaba Dental School - University of Campinas, Piracicaba, SP, 13414-903, Brazil. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Artifacts, Dental Implants, Metals. Die Arbeit nennt 45 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Finance, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior—Finance.
Abstract (Originalsprache)
Objectives
To evaluate the influence of the position of an adjacent implant, either within the endomass or exomass, and the use of the metal artifact reduction (MAR) tool on the diagnosis of peri-implant bone defects associated with the surfaces of a zirconia implant in cone-beam computed tomography (CBCT) images.
Methods
Ten bovine rib blocks were used to simulate a human mandible, each receiving 2 zirconia implants. CBCT images were acquired using the OP300 Maxio device (90 kVp, 6.3 mA, FOV 5 × 5 cm, voxel 0.125 mm) under 2 positions for the adjacent nontest implant (endomass and exomass), with and without MAR application, totaling 40 scans for the without peri-implant bone defects group. Circumferential bone defects were then created around the test implant, and 40 scans were acquired for the group with peri-implant bone defects under the same conditions. Four oral and maxillofacial radiologists individually evaluated each surface of the test implant for the presence of bone defect, using a 5-point scale. Weighted Kappa was used to assess intra- and interexaminer reproducibility. Diagnostic values were compared using multifactorial ANOVA (α = .05).
Results
In general, the position of the adjacent implant did not significantly affect the diagnostic values (P > .05). MAR application significantly increased sensitivity and area under the ROC curve for most implant surfaces, particularly the mesial and distal surfaces, regardless of adjacent implant position (P <.05). Specificity remained high under all conditions. Intra- and interexaminer agreement ranged from fair to substantial.
Conclusions
Overall, adjacent implant position—whether within the endomass or exomass—, with an interimplant distance of 5 mm, does not impact peri-implant bone defect diagnosis associated with a zirconia implant. MAR substantially improves the diagnosis of peri-implant defects on most implant surfaces, especially in the proximal area, regardless the adjacent implant position in the endomass or exomass.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Oral Diagnosis, Piracicaba Dental School - University of Campinas, Piracicaba, SP, 13414-903, Brazil.
- Department of Stomatology, Public Health, and Forensic Dentistry, School of Dentistry - University of São Paulo, Ribeirão Preto, SP, 14040-904, Brazil.
- Department of Stomatology, School of Dentistry - University of São Paulo, São Paulo, SP, 05508-000, Brazil.
ORCID-Profile: Lucas Santana-Santos, Carolina Paes Borge, Hugo Gaêta-Araujo, Matheus L Oliveira, Deborah Queiroz Freitas, Amanda Farias-Gomes
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Finance
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior—Finance
Zitieren
APA
Santana-Santos, L., Borge, C. P., Gaêta-Araujo, H., Oliveira, M. L., Freitas, D. Q., & Farias-Gomes, A. (2026). Metal artifact reduction improves cone-beam CT diagnosis of peri-implant bone defects regardless of adjacent implant position in endomass or exomass. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twag025
Vancouver
Santana-Santos L, Borge CP, Gaêta-Araujo H, Oliveira ML, Freitas DQ, Farias-Gomes A. Metal artifact reduction improves cone-beam CT diagnosis of peri-implant bone defects regardless of adjacent implant position in endomass or exomass. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twag025
BibTeX
@article{SantanaSantos2026,
author = {Santana-Santos, Lucas and Borge, Carolina Paes and Gaêta-Araujo, Hugo and Oliveira, Matheus L and Freitas, Deborah Queiroz and Farias-Gomes, Amanda},
title = {Metal artifact reduction improves cone-beam CT diagnosis of peri-implant bone defects regardless of adjacent implant position in endomass or exomass},
journal = {Dentomaxillofacial Radiology},
year = {2026},
doi = {10.1093/dmfr/twag025}
}
EUVIMED